Vincent Medical Manufacturing Co., Ltd.

Summary of Contents for inspired VHB10A

Page 1: ...Vincent Medical Manufacturing Co Ltd ...

Page 2: ......

Page 3: ...10A Humidifier Power 4 4 Alarm List 5 1 Installation of VHB10A Humidifier and Chamber 5 2 Operation 5 3 Removal of Humidification Chamber 5 4 Replace the Filter Paper for VHC25 Chamber 4 3 VHB10A Heater Plate Temperature 3 3 Product Assembly Drawing 3 4 Explanation of Symbols on the User Interface front label 2 4 Useful Information 2 5 Average Failure Free Operation Time 3 1 Attentions Cautions an...

Page 4: ...planation 10 1 Guidance and Manufacturer Declaration Electromagnetic Emissions 10 2 Guidance and Manufacturer Declaration Electromagnetic Immunity for ALL ME EQUIPMENT and ME SYSTEMS 10 3 Guidance and Manufacturer Declaration Electromagnetic Immunity for ME EQUIPMENT and ME SYSTEMS that are Not LIFE SUPPORTING 10 4 Recommended Separation Distances Between Portable and Mobile RF Communication Equip...

Page 5: ...emperature to prevent overheating Overheating can cause injury to patients and damage to the humidifier VHB10A Humidifier is suitable for use when the air flow is stable this should be 5L min In an unstable air flow environment the air tempera ture and humidity may not achieve expected values 2 1 Power Requirements 2 1 1 Operation Voltage 220 240V 2 1 2 Rated Frequency 50 60HZ 2 1 3 Power Consumpt...

Page 6: ...ening harm Cautions means the preventive measures are necessary to protect the equipments Attentions means those preventive measures procedures 2 5 Average Failure Free Operation Time is 1 000 hours 2 4 Useful Information When replacing the fuse make sure to replace with correct type and rated value fuse which is provided by the manufacturer or contact your The detailed technical information can b...

Page 7: ...y physicians The best practice established by the medical organization and or independent association should always be followed Warning Before the use of the equipment user must thoroughly understand the usage application and safety procedure information relevant to the equipment Warning Install When adding water to the humidification chamber do not let the water to exceed the maximum water level ...

Page 8: ...sufficiently long time Recommended Accessories Use humidification chamber breathing circuits and other accessories as recommended by manufacturer only Use of those accessories that are not recommended may cause the degradation of humidifier performance or affect the patient safety Danger of Electric Shock If the system is not correctly earthed electric shock may occur The system use 220 240V alter...

Page 9: ...ians for technical support and service During usage operation snould be stopped if the breathing circuit is filled with condensed water If this is the case it is recommended For the cleaning disposal method for humidification chambers breathing circuits and Ventilator or other gas supply please refers to the related and specific manuals At the end of service life dispose of the device in accordanc...

Page 10: ... It is recommended that the breathing circuit should be installed with a water trap to help remove the condensed water 3 2 Symbols The following symbols are used in the product their meanings are explained below 0197 Electrial Shock Hazard Refer to qualified service personal Altemating Current IPX1 Dip proof protection to IPX1 Refer to instruction manual booklet Protective earth ground OFF power O...

Page 11: ...t assembly Main PCB assembly Cord clamp Screw Gasket Nut Registered information label Bracket Power cord assembly Transpatent rubber foot Power switch Bottom case Micro switch assembly Power connector wire assembly Power supply PCB assembly Heater plate temp detector assembly Knob Front label Button seat Chamber lock Spring Transformer Power label Internal bracket Inside Bracket Stretch gasket Was...

Page 12: ...ously 4 2 VHB10A Humidifier Power The maximum power is 100VA 15 4 3 VHB10A Heater Plate Temperature VHB10A Humidifier s temperature can be adjusted in the range of 45 80 Normal Working Condition A thermostat is installed to limit the temperature of the heater plate to be less than 95 7 If the electric circuit is faulty the alarm buzzer will turn on if heater 4 4 Alarm list The alarms above are tec...

Page 13: ...dd sterilized water into the chamber until the water level reach the maximum water level line indicated on the chamber Please read the chamber instructions to understand the details related to the use of chamber 5 1 4 Slide the chamber into the humidifier slot then use the chamber to press down the button at the top of the humidifier until the button springs back up indicating the chamber is firml...

Page 14: ...wer cord to the suitable alternating current socket with earthing the power source should have rated power as specified as per this instruction Now the humidifier and chamber is ready Before use if you have any questions on how to use the humidifier properly please contact Vincent Medical Mfg Co Ltd or your local supplier VHB10A is intended for use in conjunction with gas supply such as iVent 201o...

Page 15: ... switch on the humidifier make sure air is flowing through the chamber and the breathing circuit 5 Do not use the device with the flammable anesthetic gases as this may cause explosion 6 Use the device only with the Vincent Medical Manufacturing Co Ltd humidification chambers and breathing circuits as recommended by the 5 2 Operation 5 2 1 Ensure the power cord of the humidifier is connected to an...

Page 16: ... 2 5 Refer to the relevant section of this manual for the humidifier control monitor alarm and setting information 5 2 6 5 2 7 Periodic check on the gas temperature at the patient end of inspira tory limb must be done to ensure the temperature is appropriate and prescribed to ensure patient safety 5 2 8 If the gas flow is interrupted the humidifier must be first turned off Once the gas flow has re...

Page 17: ...y cause explosion 7 If the air flow is interrupted please switch off the humidifier 8 Do not add water beyond the maximum level as indicated on the humidification chamber Use the the humidification chamber in accordance to the instructions in the operator manual 5 3 Removal of humidification chamber 5 3 1 Turn off the humidifier through the power switch at the side of the equipment 5 3 2 Disconnec...

Page 18: ...situation happens repeatedly please contact qualified technician for technical support 7 1 Basic inspection Everytime before use inspection on the whole system is recommend ed Visual inspection on humidifier heater plate accessories in use and power cord should be performed by qualified and trained technicians clinicians physicians at least every month If necessary the damaged parts should be repl...

Page 19: ...r or Vincent Medical Manufacturing Co Ltd 2 Micro switch failure 1 The overheating protection switch hermostat is off 3 Storage and transportation no power The heating indicator is flashing in red alarm is sounding 2 Malfunction of heater plate probe 2 Contact local supplier or Vincent Medical Manufacturing Co Ltd 1 The humidification chamber is not installed correctly so the microswitch under the...

Page 20: ...2018 07 09 A10022177 REV B Page 18 of 24 10A length of 10A compatible A ...

Page 21: ...2018 07 09 A10022177 REV B Page 19 of 24 A A A RF emissions CLSPR 11 RF emissions CLSPR 11 Harmonic emissions IEC 61000 3 2 IEC 61000 3 3 IEC 61000 4 2 IEC60601 1 1998 ...

Page 22: ...2018 07 09 A10022177 REV B Page 20 of 24 A A Surge IEC 61000 4 5 Electrostatic fast transient burst IEC 61000 4 4 IEC 61000 4 11 ...

Page 23: ...2018 07 09 A10022177 REV B Page 21 of 24 A IEC 61000 4 6 IEC 61000 4 3 IEC61000 1 1988 IEC 61000 4 8 ...

Page 24: ...2018 07 09 A10022177 REV B Page 22 of 24 A A A ...

Page 25: ...2018 07 09 A10022177 REV B Page 23 of 24 A A A A ...

Page 26: ...10A so they should be away from VHB10A when VHB10A is in operation 3 The power cord is 3 0m in length non shielding please do not replace lengthen shorten the power cord otherwise will cause the equipment to increase its emission or decrease its immunity 4 VHB10A should not be used nearby stack up with other equipments if VHB10A must be used nearby stack up with other equipments then user must obs...

Page 27: ......

Page 28: ...treet Hung Hom Kowloon Hong Kong Tel 852 2186 1010 Fax 852 2765 8428 E mail service inspired medical com Website www inspired medical com Vincent Medical Manufacturing Co Ltd Flat RM B2 7 F Hang Fung Industrial Building Phase 2 2G Hok Yuen Street Hung Hom Kowloon Hong Kong Tel 852 2365 5688 Fax 852 2765 8428 ...