2018-07-09 A10022177 REV.B Page: 10 of 24
3.4 Explanation of the symbols on user interface (front label)
VHB10A User interface
4.1 Temperature Setting Range
The range for heater plate
temperature setting:
45
℃
~80
℃
4.1.2 Heating Indicator
When the humidification
chamber is installed to the humidifier, the indicator on the right will
Iight on in green continuously to indicate heating is in process.
4.1.1 Power Indicator
When power switch is on,
the left power indicator will
Iight on in green continuously.
4.2 VHB10A Humidifier Power
The maximum power is 100VA + 15%
4.3 VHB10A Heater Plate Temperature
VHB10A Humidifier's temperature can be adjusted in the range of
45~ 80
℃
(Normal Working Condition); A thermostat is installed to
limit the temperature of the heater plate to be less than (95±7)
℃
,
If the electric circuit is faulty,( the
alarm buzzer will turn on if heater
℃
.
4.4 Alarm list
The alarms above are technical alarms. The heating indicator will be
flashing in red, and the alarm sound level is of a range of 50-60dB.
Alarm priority # Alarm description Mute
Alarm delay
NO
<10s
<10s
NO
High
Malfunction of
heater plate probe
The overheating
protection switch
(
Thermostat
)
is off
High
plate higher than(95±7)