19616-001 Rev. D
DiagnosticPRO
Edge
Film Digitizer Installation Guide
38
Appendix:
Electromagnetic guidance
Caution: Medical electrical equipment.
EMC (Electro Magnetic Compatibility) must be considered before any medical electrical equipment is installed or
put into service. Follow the information in the accompanying documentation when installing and operating the
DiagnosticPRO
Edge
Digitizer.
Caution: Portable or mobile RF communication equipment can effect Medical Electrical equipment.
Caution: Using the Digitizer adjacent to or stacked with other equipment may cause interference between the
equipment. Before utilizing stacked or adjacent equipment, verify proper functionality of all equipment in the
actual configuration in which it will operate.
Caution: Connecting the Digitizer to equipment that is not rated CISPR 11 class A or class B may alter the
electromagnetic characteristics.
Caution: In a domestic environment this product may cause radio interference in which case the user may be
required to take adequate measures.
Guidance and manufacturer’s declaration –
electromagnetic equipment
Table 201
The DiagnosticPRO
Edge
Digitizer is intended for use in the electromagnetic environment specified below. The customer
or end user of the DiagnosticPRO
Edge
Digitizer should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF Emissions
CISPR11
Group 1
The DiagnosticPRO
Edge
Digitizer uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF Emissions, CISPR11
Class A
The DiagnosticPRO
Edge
Digitizer is suitable for use in all
establishments other than domestic and those directly connected to
the public low voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
flicker emissions
IEC 61000-3-3
Complies