17
9.4.1
9.4.1
9.4.1
9.4.1
100mV
100mV
100mV
100mV
Range
Range
Range
Range
calibration
calibration
calibration
calibration
1) The calibration wiring is shown in the following diagram:
2) Press the key
〔
INPUT/OUTPUT
INPUT/OUTPUT
INPUT/OUTPUT
INPUT/OUTPUT
〕
to enter the calibrator in a
state of 100mV input calibration when the display indicates the
symbols ‘INPUT’, ‘CAL 0’, ‘ON’, and ‘100.00mV’.
3) Set the standard source to a corresponding range.
4) Set the output of standard source to the indication of the
calibrator. With the output stabilized, press the key
〔℃
///
/
℉〕
and
the display will flash, denoting that the calibrated point has
been stored.
10
10
10
10
...
.
Points
Points
Points
Points
for
for
for
for
Attention
Attention
Attention
Attention
to
to
to
to
Use
Use
Use
Use
of
of
of
of
Operation
Operation
Operation
Operation
Instruction
Instruction
Instruction
Instruction
The present operation instruction is subject to change without
notice.
The content of the operation instruction is regarded as correct.
Whenever any user finds its mistakes, omission, etc., he or she
is requested to make contact with the manufacturer.
The manufacturer is not liable for any accident and hazard
arising from any maloperation.
The functions described in this operation instruction should not
be used as grounds to apply this product to a particular purpose.
1
DCV/100mV
FS:100mV
TRIG
OUT
Lo
OUTPUT
INPUT
SCOPE
Hi
NORMAL
AUX
30V
MAX
30V
MAX
Lo
Lo
Hi
Hi
standard source
(
5520A
)