5.  Directions for Use

1.  Press the power        button on front of unit. The LCD is activated to 

show all indicators for 2 seconds.

2.  When the screen shows “00”, a beep is heard and the thermometer is 

ready for measurement.

3.  Place the probe gently and flush onto the area approximately 1/2 inch 

above the center of the eyebrow. If the eyebrow area is covered with 
hair, sweat or dirt, please clean the area beforehand to improve the 
reading accuracy.

4.  Press the power        button once. Move the thermometer steadily 

from 1/2 inch above the center of the eyebrow to 1/2 inch above the 
tail of the eyebrow area and back to the center of the forehead. A long 
beep will sound after 3 seconds to signal measurement complete. If 
you do not reach the tail of the eyebrow before the long beep is heard, 
repeat the measurement, moving the thermometer a little quicker.

5.  Read the recorded temperature from the LCD display.

 

Repeat measurement if necessary - repeat step 3 above and follow 
instructions.

NOTE: The unit will shut off automatically after 30 seconds.

(3)

(1)

(2)

(4)

(5)

To avoid an inaccurate measurement, note the following:

 

•   Patients should stay in steady - state room condition for at least 30 minutes.
•   Don’t remove the measurement device from the measuring area before hearing the

 

final beep.

•   Don’t take a measurement while or immediately after nursing a baby.
•   Don’t use the forehead thermometer in high humidity environments.
•   Patients should not drink, eat, or be physically active before/while taking the measurement.
•   Before placing the thermometer sensor onto the measurement area, remove dirt, hair,

 

or sweat.

•   Use an alcohol swab to carefully clean the sensor and wait for 5 minutes before taking a 

measurement on another patient.

•   Always take the temperature in the same location, since the temperature readings may 

vary from different locations.

•  In the following situations it is recommended that three temperatures in the same location 

be taken and the highest one taken as the reading:
1)  Newborn infants in the first 100 days.
2)  Children under three years of age with a compromised immune system and for whom the 

presence or absence of fever is critical.

3)  When the user is learning how to use the thermometer for the first time until he/she has 

familiarised himself/herself with the instrument and obtains consistent readings.

7.  MemoryTracker™ Mode

Use the following steps to recall the last 8 temperature measurements.

Display

Description

Press and release the     button to recall the last 
reading. The number 1 will appear showing the last 
saved reading. Press the       button again to scroll 
through the other saved readings in memory.

M

To exit memory mode and turn the thermometer off, press and hold the     button for 3 seconds. 
To exit memory mode and take another temperature, press the power         button.

M

6.  Control Displays and Symbols

LCD Display

Display Meaning

Description

Self check ensures 
proper functioning

Press the power button to turn 
on the unit, all indicators will 
be shown for 2 seconds.

Ready to measure

The unit is ready for the 
measurement, the screen 
displays steady “00”.

Measurement complete

The reading will be shown on 
the LCD display.

Low battery indication

When the unit is turned on, the 
battery icon will keep flashing 
to remind the user to replace 
the batteries.

M