CONFIGURATION
After connecting the fiber to be tested to the optical
connector, you must select the Nano-OSA module (see
the Base Unit user manual).
To configure the Base Unit in preparation for an OSA test
on a fiber, press the
SETUP
button. Two configuration
procedures are available:
•
“Auto-set mode”
In this case the parameters are selected automatically.
•
“Manual configuration”
In this case you may define your own configuration.
Auto-set mode
◮
Press the Auto-set softkey.
Following configuration will automatically be set.
Parameter
Setting
Acquisition settings
Sweep
Single
Measurements setting
Channel detection
All
Signal threshold
Auto
Splitter compensation
No
Results screen settings
Grid
conventional
Channel Width
Auto
Alarms
No
Unit of x-axis
last setting used (THz or nm)
In case of any parameter setting the
Test Auto WDM
softkey appears.