Chapter 1
Optical Spectrum Measurement
Configuration
MTS / T-BERD 5800 • COSA-4055 • User Manual 2016.01
11
Acquisition
parameters
Sweep
Number of sweeps
–
In Statistics mode: between 2 and 1000
–
In Drift mode: between 2 and 10000
Wait Period
The Wait Period parameter allows to enter a wait period prior to the
measurement start (in Statistics and Drift mode only).
s
Continuous
There will be a measurement with refreshment of
the trace and real time display of results.
Single
There will be one single measurement and the
display of its results.
Statistics
In this mode, the number of samples concerned by
statistics must be entered (next parameter).
Drift
For measuring power and wavelength over time.
The number of sweeps and the latency between
the sweeps need to be set.
NOTE
For Drift measurements a Grid needs to be defined and the Channel
Detection parameter is set to Grid.
Up to ...
Increments of ...
1 minute
5 seconds
10 minutes
1 minute
60 minutes
5 minutes
24 hours
1 hour