14
(3)
Press the right middle finger on the back electrode.
(4)
Press the left electrode to the left palm.
6.
Once the device detects stable waveform, it will automatically start
the measurement. The countdown bar moves from left to right.
7.
When the bar is fully filled, the device will analysis your data, and
then show the measurement result.
Please refer to
Section 2.4
to understand the result screen. Daily Check
provides the trending graph of heart rate, SpO
2
and blood pressure or
change in percentage of blood pressure. To view the trend, tap the
button, then select one record, and then tap the
button. For details,
please refer to
Section 6.1
.
Set reference
To track the blood pressure change, you need to set one measurement as
the reference. For a given user, if the reference has not been set, then
when finishing the Daily Check, tap the
icon to set current
measurement as the reference. If the reference has already been set, tap
the
icon again will set the current measurement as a new reference.
Warnings and Cautionary Advices
For a given user, it is suggested to set a new reference every three
months.
BP Calibration
To get blood pressure readings, this device should be calibrated by a
doctor with a traditional cuff blood pressure (BP) meter. Because of
individual differences, each user must make his/her own calibration
before using Daily Check to measure or track the blood pressure. The
calibration should be performed when the user is under calm status.
To calibrate with a cuff BP meter, follow the steps as below.
1.
Find a traditional cuff BP meter.
2.
Sit down and stay calm.
3.
Place the cuff on you left arm according to instructions.
4.
Pick up the Checkme monitor, press Home button to enter Main