3–27
Chapter 3 Maintenance and Troubleshooting
F6 Sample pump
When checked sample pump is turned on
F7 Pressure Zero
When checked,
zeros all pressure transducers and checks the shunt box so that the
system is referenced to ambient
F10 Reset
Clears all checkmarks
D/A D
Allows user to output a voltage through the pod box
D/A E
Allows user to output a voltage through the pod box
D/A F
Allows user to output a voltage through the pod box
D/A G
Allows user to output a voltage through the pod box
Ambient Module
Temperature
Temperature Reading taken from the ambient module connected to an USB port on the
Computer. This temperature overrides the module temperature
Humidity
Humidity Reading taken from the ambient module connected to an USB port on the
Computer.
Sol 1-13 DLCO, Alarm, Fan
A Insp/Cal Sample
Switches between inspire port on back of module (unchecked) and Calibration port on
the front of the module (checked)
B Dil/Cal Sample
Switches between the mixing chamber (unchecked) and the calibration port on the
front of the module (checked)
C Hi/Lo Sample
Used for calibration. Used to determine the pressure effects on the O2 sensor
D Venturi
Forces evacuation of the balloon valves
E Gas Port
Inflates the left balloon valve when checked
F Spir Port
Inflates the right balloon valve when checked
G Valve Enable
Allows for E and F to work, when balloons inflated and G is disabled the balloons
should stay inflated
H Cal 2
Turns on cal 2 gas
I Cal 1
Turns on cal 1 gas
J DL/O2 Cal
If checked and DLCO (L) is checked, then DLCO gas will be sent through the
calibration circuit. If checked and DLCO (L) is unchecked then O2 gas will be sent
through the calibration circuit.
K Purge
Flushes the breathing hose (22 mm) with either O2 or DLCO gas (if L is checked)
L DLCO
Unchecked supplies oxygen as the test gas. Checked supplies DLCO as the test gas.
M Alarm
Turns on the mixing chamber alarm
N Fan
Turns on the mixing chamber fan
Summary of Contents for Vmax Encore
Page 1: ...L2707 Rev A 2005 VIASYS Respiratory Care Inc Technical Reference Manual Vmax Encore ...
Page 4: ...Vmax Encore Technical Reference Manual iv L2707 ...
Page 6: ...Vmax Encore Technical Reference Manual vi L2707 ...
Page 26: ...Vmax Encore Technical Reference Manual 1 20 L2707 ...
Page 54: ...Vmax Encore Technical Reference Manual 2 28 L2707 ...