Chapter 2 System Description
2–20
L2707
Technical Reference Manual
Vmax Encore
turn to green when the mean flow rate of a stroke falls within the adjacent range.
If you complete 15 strokes, or the three-minute clock reaches zero, before you turn on the
required number of green segments, the following message will be displayed:
If the required number of bar graph segments is turned to green before you complete the 15
strokes, or before the three-minute timer reaches zero, the
Calibration Bar Graph
screen is
replaced by the
Calibration Verification
window. The following message will then be dis-
played:
This message is displayed either automatically or manually, as designated in the
Flow Vol-
ume Setup Calibration
dialog box. Refer to the
Vmax Reference Manual
for a description
of this dialog box. You can now proceed to the next step.
6.
In the
Verification
window, perform five full inspiratory and full expiratory strokes. Perform
these strokes at the ATS (American Thoracic Society) recommended flow rates.
Four of the five strokes are displayed and should appear as follows:
•
One stroke (inspiratory and expiratory) should reach the lowest dotted line (0.5 LPS).
•
One stroke (inspiratory and expiratory) should reach the highest dotted line (3.0
LPS).
•
One stroke should be halfway between the dotted lines (1.5 LPS).
•
One stroke (the fourth stroke) should represent a peak-flow rate near 12 LPS (8 LPS
minimum). This stroke should be created without “banging” the piston at either end of
the stroke.
Minimum Calibration
Requirements have not
been met.
F1 to repeat calibration.
This message will be given if:
•
There are less than three
green segments for the
inspiratory strokes (in any
combination).
•
There are less than three
green segments for the expi-
ratory strokes, and the sec-
ond and third segments are
not green.
Select
F1
to repeat the calibration;
select
Esc
to terminate the
calibration and return to the Mass
Flow Sensor Calibration screen.
Minimum Calibration
Requirements have been
met.
F1 to repeat the calibration.
Select
F1
to repeat the calibration;
select
F2
to accept the calibration.
Summary of Contents for Vmax Encore
Page 1: ...L2707 Rev A 2005 VIASYS Respiratory Care Inc Technical Reference Manual Vmax Encore ...
Page 4: ...Vmax Encore Technical Reference Manual iv L2707 ...
Page 6: ...Vmax Encore Technical Reference Manual vi L2707 ...
Page 26: ...Vmax Encore Technical Reference Manual 1 20 L2707 ...
Page 54: ...Vmax Encore Technical Reference Manual 2 28 L2707 ...