Chapter 3 Maintenance and Troubleshooting
Figure 3-1
Calibration Syringe Adjustment/Attachment for FRC Verify
Note:
Do
not
use the flexible Calibration Adapter or any mixing chamber hose to connect
the syringe to the Mass Flow Sensor; use the plastic FRC Adapter.
8.
Stroke the syringe two times to fully flush the breathing valve with room air.
Reposition the
piston shaft at the etched line
.
9.
Select
F1
to start the measurement procedure. Begin stroking the piston slowly, pulling the
piston shaft all the way out, then pushing it back in to the etched line. This is the stroke vol-
ume for the FRC Verification (about 1/3 the full stroke volume). Try to be as accurate as pos-
sible, pulling the piston all the way out, then pushing it back in to the etched line.
Note:
During the FRC Verification procedure, the Lung Volumes Test screen is identical to
screen displayed during a regular patient test, except for the Text Display at the top of
the screen.
The Text Display shows the predicted FRC volume for the calibration syringe, the
accumulating measured FRC volume, and the ratio of the measured volume to the
predicted volume expressed as a percent.
0511822
Summary of Contents for Vmax Encore
Page 1: ...L2707 Rev A 2005 VIASYS Respiratory Care Inc Technical Reference Manual Vmax Encore ...
Page 4: ...Vmax Encore Technical Reference Manual iv L2707 ...
Page 6: ...Vmax Encore Technical Reference Manual vi L2707 ...
Page 26: ...Vmax Encore Technical Reference Manual 1 20 L2707 ...
Page 54: ...Vmax Encore Technical Reference Manual 2 28 L2707 ...