38
Bear
Cub
750PSV
NOTE
If problems are incurred in obtaining the calibration performance, refer to
Troubleshooting (
Chapter
6) or contact your dealer or VIASYS Healthcare Service
Facility.
Pressure Transducers and Flow Control Valve
Calibration
Turn OFF the ventilator and disconnect tubing to the Pressure Transducers on the
Control PCB. Press and hold the diagnostic switch on the Control PCB while turning
the ventilator ON to enter Transducers and Flow Calibration routines. Throughout
calibration, the TEST key is for advancing the calibration number; VISUAL RESET is
for backing to previous calibration, and the MANUAL BREATH key is for recording
calibration data. The calibration number and message are displayed in the Breath
Rate and Monitored Time/I:E Ratio windows respectively. Data (when applicable) is
displayed in the Monitored Pressure window. Refer to the following (See table 4-1)
to calibrate the transducers on the Control PCB:
Table 4-1: Control PCB Transducer Calibration
Transducer
Calibration Number
Message
Input Static Pressure
Machine (port B)
d00
nP00
0 cmH2O
d01
nP40
40 ± 0.2 cmH2O
Differential (port B)
d02
dP00
0 cmH2O
d03
dP40
40 ± 0.2 cmH2O
Proximal (port B)
d04
pP00
0 cmH2O
d05
pP40
40 ± 0.2 cmH2O
O2 (A2)
d06
0P00
0 PSIG
d07
0P40
40 ± 0.2 PSIG
AIR (A1)
d08
AP00
0 PSIG
d09
AP40
40 ± 0.2 PSIG
NOTE
The 40 cm/psi settings were 70 cm/psi prior to monitor firmware 511-02803-00.
When calibrating the pressure transducers, both pressure points (i.e. 0 and 40
cm/psi) must be read and the 0 cm/psi reading must be obtained first.
1.
Check the accuracy of the five pressure transducers per table 4.2. The
window shows the pressure readings monitor pressure.
2.
Press the TEST key to advance the calibration number to the Display Set
Altitude Test, then press the MANUAL BREATH key until the number in the
PRESSURE window indicates the altitude at which the Flow Control valve
calibration is to be performed (expressed in units of 100 ft). Then advance to
the next test number and power down.
3.
Re-connect tubing to the transducers on the Control PCB; apply Air and O2
(approximately 50 psi) to UUT. Use an RT-200 to measure flow at the TO
L2346
Revision B
June 2004
Summary of Contents for Bear Cub 750PSV
Page 6: ...6 Bear Cub 750PSV L2346 Revision B June 2004 ...
Page 12: ...12 Bear Cub 750PSV L2346 Revision B June 2004 ...
Page 24: ...24 Bear Cub 750PSV Figure 2 2 Pneumatics System Schematic L2346 Revision B June 2004 ...
Page 36: ...36 Bear Cub 750PSV L2346 Revision B June 2004 ...
Page 44: ...44 Bear Cub 750PSV L2346 Revision B June 2004 ...
Page 46: ...46 Bear Cub 750PSV Figure 5 2 Power Supply Module Assembly 2 of 2 L2346 Revision B June 2004 ...
Page 47: ...Service Manual 47 Figure 5 3 Schematic Diagram Pneumatic 750PSV L2346 Revision B June 2004 ...
Page 48: ...48 Bear Cub 750PSV Figure 5 4 Power Supply PCB Assembly 1 of 3 L2346 Revision B June 2004 ...
Page 49: ...Service Manual 49 Figure 5 5 Power Supply PCB Assembly 2 of 3 L2346 Revision B June 2004 ...
Page 50: ...50 Bear Cub 750PSV Figure 5 6 Power Supply PCB Assembly 3 of 3 L2346 Revision B June 2004 ...
Page 51: ...Service Manual 51 Figure 5 7 Pneumatic Tubing Diagram L2346 Revision B June 2004 ...
Page 52: ...52 Bear Cub 750PSV Figure 5 8 Display PCB 1 of 3 L2346 Revision B June 2004 ...
Page 53: ...Service Manual 53 Figure 5 9 Display PCB 2 of 3 L2346 Revision B June 2004 ...
Page 54: ...54 Bear Cub 750PSV Figure 5 10 Display PCB 3 of 3 L2346 Revision B June 2004 ...
Page 55: ...Service Manual 55 Figure 5 11 Control PCB Assembly L2346 Revision B June 2004 ...
Page 56: ...56 Bear Cub 750PSV Figure 5 12 Wiring Diagram L2346 Revision B June 2004 ...
Page 57: ...Service Manual 57 Figure 5 13 Power Supply Schematic L2346 Revision B June 2004 ...
Page 58: ...58 Bear Cub 750PSV Figure 5 14 Control Schematic PCB 1 of 5 L2346 Revision B June 2004 ...
Page 59: ...Service Manual 59 Figure 5 15 Control Schematic PCB 2 of 5 L2346 Revision B June 2004 ...
Page 60: ...60 Bear Cub 750PSV Figure 5 16 Control Schematic PCB 3 of 5 L2346 Revision B June 2004 ...
Page 61: ...Service Manual 61 Figure 5 17 Control Schematic PCB 4 of 5 L2346 Revision B June 2004 ...
Page 62: ...62 Bear Cub 750PSV Figure 5 18 Control Schematic PCB 5 of 5 L2346 Revision B June 2004 ...
Page 63: ...Service Manual 63 Figure 5 19 Display Schematic PCB 1 of 4 L2346 Revision B June 2004 ...
Page 64: ...64 Bear Cub 750PSV Figure 5 20 Display Schematic PCB 2 of 4 L2346 Revision B June 2004 ...
Page 65: ...Service Manual 65 Figure 5 21 Display Schematic PCB 3 of 4 L2346 Revision B June 2004 ...
Page 66: ...66 Bear Cub 750PSV Figure 5 22 Display Schematic PCB 4 of 4 L2346 Revision B June 2004 ...
Page 68: ...68 Bear Cub 750PSV L2346 Revision B June 2004 ...
Page 96: ...96 Bear Cub 750PSV Figure 11 1 Repacking Procedure L2346 Revision B June 2004 ...