REPLACING/MAINTAINING IMMERSION UV-C LAMP
The lamp must be replaced after 8.000 or 16.000 (amalgam) operating hours and the quartz
glass must be cleaned.
Drawing 1
1. For your own safety, put on the supplied UV safety goggles (A)
2. Wear the supplied gloves (B)
3. De-energize the appliance and wait at least 10 minutes
4. Carefully take the Immersion UV-C out of the appliance/machine
5. Loosen the nut (I) from the screw part (H)
6. Carefully pull the lamp (C) off the lamp fitting (F)
7. Take a new UV lamp (C) out of the packaging
8. Carefully slide the lamp (C) into the quartz glass (D)
9. Place the lamp fitting (F) on the lamp (C)
10. If necessary, clean the quartz glass with a soft cloth and a non-aggressive cleaning agent.
When the quartz glass is totally dry you can check for lime scale deposits. Remove any lime
scale deposits with appropriate lime scale removers (i.e. cleaning vinegar).
11. Slide the lamp fitting into the quartz glass and hand-tighten the nut (I) onto the screw part
(H). This will ensure that the O-ring (J) is locked in and that the connection is therefore
watertight.
12. The UV-C appliance is now ready for use
TECHNICAL DATA
Model
40W
75W
40W Amalgam
80W Amalgam
Article number
3100860
3100862
3100864
3100866
T5 lamp
40 Watt
75 Watt
40 Watt
80 Watt
Lamp life
8000 hours
8000 hours
16000 hours
16000 hours
Supply voltage
230V AC
50/60Hz
230V AC
50/60Hz
230V AC
50/60Hz
230V AC
50/60Hz
Length of
lum
quartz glass
85 cm
85 cm
55 cm
85 cm