VC series bench scales Technical Manual
- 19 -
8.4.5
Output control setting
Command
Contents
Parameter
Address
・
Value
Response
C1 C2
Code
O
0
4FH 30H
No output
(
Able to input command
)
Nil
A00
:
Proper
completion
O
1
4FH 31H Continuous output
O
2
4FH 32H
Continuous output when stable
(
No output when unstable
)
O
3
4FH 33H
One output each time
pressings a key
O
4
4FH 34H Auto output
O
5
4FH 35H
One output when stable
(
Stop when unstable
)
O
6
4FH 36H
One output when stable
(
Continuous output when
unstable
)
O
7
4FH 37H
One output after pressing a
key when stable
O
8
4FH 38H Instant one output
O
9
4FH 39H One output after stable
Note1:
Output control by
“
O0
” to “
O7
”and output control by function setting of a
scale work same. With regard to a supplement and notice of each operation, refer to
“ 7. Function settings of a scale”
“
O8,O9
”command means the data demand command to a scale.
Note2:
When the set each condition works properly, the data , which is set in the
function setting of a scale (41. da. ), is transmitted.
Note3:
After carrying out
“
O0
” to “
O9
” command, the condition is kept until next
command is inputted. However, in the case that it is tuned off and on again, output
control becomes the initial value (Function set value).
8.4.6
Data output demands
Command
Contents
Parameter
Response
C1
C2
code
Addres
s
Value
W
1
57H
31H
Weight value output
Nil
Nil
Weight value data
C
1
43H
31H
Quantity value output
Quantity value data
C
2
43H
32H
Unit weight value output
Unit weight value
data
C
3
43H
33H
Total value output
Total value data
L
1
4CH
31H
Lower limit value output
Lower limit value data
L
2
4CH
32H
Upper limit value output
Upper limit value data