VC series bench scales Technical Manual
- 12 -
P2 d M
Dual ra
Deci
To set decimal points.
Options:
0 , 0.0 , 0.00 , 0.000 , 0.0000
inc
Di v 1
To select first division
Options: 1 , 2 , 5 , 10 , 20 , 50
Di v 2
To select second division
Options: 1 , 2 , 5 , 10 , 20 , 50
Cap
Cap 1
To select first capacity
Cap 2
To select second capacity
Cal
Calibration
Dual in
To select dual interval
Note:
First interval will active in CAP 1
Second interval will active in CAP 2
Deci
To set decimal points
inc
Di v 1
To select first division
Options: 1 , 2 , 5 , 10 , 20 , 50
Di v 2
To select second division
Options: 1 , 2 , 5 , 10 , 20 , 50
Cap
Cap 1
To select first capacity
Cap 2
To select second capacity
Cal
Calibration
SECTION 7. CALIBRATION
Select technical parameter CAL mode and select desired setting Linear (linear
calibration) and nonLin (Normal Calibration)
7.1 Normal Calibration: nonLin
Enter the function by pressing
, display will be shown
Make sure there are no loads on the platform and wait few seconds for stable
indicator on.
Display will be show
Load the standard calibration mass weight on the platform and wait
few seconds for display stability.
Display will be show
After the calibration the display will start a self test. Remove the load from platform
during the test. Display will come to weighing mode automatically.
unload
load
Pass