APPENDIX 1:
ERROR MESSAGE
1. Error message
Err. Msg.
Problem
Err2
Initial zero point over ±10% (take 10% as the benchmark)
Err3
Over or below A/D resolution range
FFFFF
Err4 EEPROM
Chksum error
Err5
Overload (max. capacity
+9e)
Err6
Wrong calibration weight
Battery
symbol
Low battery
2. Troubleshooting
When Err.
Msg.
Troubleshooting
Power on
Err2
Check and remove any object from weighing pan or malfunction of LOAD CELL
Power on
Err3
Check if A/D or LOAD CELL malfunctions
Power on
Err4
Beep alert. Switch power off and power on again, or perform calibration
Power on
Battery symbol
Charge the battery with power on
Operating at
normal
weighing
mode
Err5
Checking if the weighing object is over the max. ca9e
Calibrating
Err6
Change weight
Instruktion-LPWN Rev 1.2005
Utskrift 2005-12-01
Användarmanual
Vetek AB
Box 79, Industriv. 3 760 40 Väddö
Tel 0176 208 920
Fax 0176 208 929
e-mail: info@vetek.se
www. vetek.se
Sida
9