V. KEYPAD 

 

1.“PRINT” 

Press PRINT key, RS-232 will print data, when negative value is shown on the screen, the printing 

function is prohibited. 

 

2.“MODE” 

Press MODE key for selecting weighing unit. 

 

3.“TARE” 

Press TARE key for deduction of current weight on the weighing pan. After removing the weight on 

the weighing pan, press TARE key to cancel the original tare. 

Note: This operation can only be performed in the stable mode. 

 

4.“ZERO” 

Press ZERO key and the screen will show zero 

0.0000

, press ZERO key in the tare mode, the tare will 

be cancelled; if the weight on the weighing pan is within the zero adjustment range, the scale will be 
reset to zero. 

 

5. “ON/OFF” 

Power switch. 

 

 
Instruktion-LPWN Rev 1.2005 

Utskrift 2005-12-01 

 

 

 Användarmanual    

 

 

     

 

Vetek AB 

Box 79, Industriv. 3  760 40 Väddö 

 

Tel  0176 208 920 

Fax 0176 208 929  

                     

e-mail: [email protected]

                                 

     

  

www. vetek.se

                                 

      

Sida  

5

 

Summary of Contents for LPWN-1530

Page 1: ...ntrum av den inre ringen c Vågen bör alltid användas i en miljö som är så fri som möjligt från temperaturvariationer damm vibrationer drag fukt eller i korrosiv miljö Dessa faktorer påverkar hållbarhet och tillförlitlighet d För kontinuerlig drift anslut vågen till nätet med nätadaptern Vi rekommenderar att inte maskiner tex borrmaskiner används på samma säkring Vi rekommenderar att vågen ansluts ...

Page 2: ...behållaren med innehåll visas behållarens vikt så som negativt värde Tryck TARE för att återställa vågen Att räkna småsaker Detta är en våg för vanlig vägning men den har även en enkel räknefunktion Räkna enheter Vågen är utrustad med genomsnitt beräkning Denna funktion kompenserar för mindre viktskillnader i provvikterna Provvikterna bör väga minst 0 8 x vågensgradering Är vikterna för små för at...

Page 3: ...3 Easy to read LCD display 4 Counting function is applicable with backlight function 5 Tare range is unlimited 6 The case shell is made of ABS shockproof plastics stainless steel weighing pan is used for long term operation III ASSEMBLY A Models LPWN 1530 LPWN 7515 LPWN 1530K 1 Remove the protection copper screw 2 and then screw out the protection copper screw 1 2 Align the three hole aluminum col...

Page 4: ...on the LCD display charge the battery with AC power cord plug in the indicator of charge will light up in red when it becomes green means charge completed It takes about 8 hours to full charge the battery 2 Install the equipment on a level and stable surface 3 Do not install the equipment near the air conditioning or a vibrating machine 4 Install the equipment in an environment with steady tempera...

Page 5: ...E key to cancel the original tare Note This operation can only be performed in the stable mode 4 ZERO Press ZERO key and the screen will show zero 0 0000 press ZERO key in the tare mode the tare will be cancelled if the weight on the weighing pan is within the zero adjustment range the scale will be reset to zero 5 ON OFF Power switch Instruktion LPWN Rev 1 2005 Utskrift 2005 12 01 Användarmanual ...

Page 6: ... 1 2 3 4 Page 5 Item 4 Auto Auto power off OFF 5 10 30 60 Page 5 Item 5 Ligh Backlight OFF ON AUT Page 5 Item 6 Lit Zero point range 0 1 2 3 4 5 Page 5 Item 7 Baud RS 232 baud rate 2400 4800 9600 Page 5 Item 8 rS232 Printing mode Co St Pr Page 5 Item 9 Print External device PC SH EZ Page 5 Item 10 Stabl Printing stability Yes No Page 5 Item 11 Instruktion LPWN Rev 1 2005 Utskrift 2005 12 01 Använd...

Page 7: ...ime 6 Press MODE key once again the LCD display will show Ligh which means the scale enters into backlight setting mode Press TARE key to select setting if the display shows OFF it means the backlight is off ON means the backlight is on Auto means auto backlight is on 7 Press MODE key once again the LCD display will show Lit press TARE key d1 will appear on the display press TARE key again to sele...

Page 8: ...ARE key until On2 appears put weight of 2 3 full load on the weighing pan press TARE again On3 will appear on the display press TARE after putting weight of full load on the weighing pan and wait until PASS appears 3 Take off the weight press ZERO key to return to weighing mode B SINGLE POINT CALIBRATION 1 Power on while holding down MODE key to enter into setting mode the display shows CAL 2 Pres...

Page 9: ...object from weighing pan or malfunction of LOAD CELL Power on Err3 Check if A D or LOAD CELL malfunctions Power on Err4 Beep alert Switch power off and power on again or perform calibration Power on Battery symbol Charge the battery with power on Operating at normal weighing mode Err5 Checking if the weighing object is over the max capacity 9e Calibrating Err6 Change weight Instruktion LPWN Rev 1 ...

Page 10: ... N W 1 1375kg T W 0 0180kg G W 1 1555kg 2 g G N W g C R LF weight Example N W 1137 5g T W 18 0g G W 1155 5g 3 lb G N W l b C R LF weight Example N W 2 508lb T W 0 040lb G W 2 548lb 4 lb oz G N W l b o z C R L F weight lb Weight oz Example Instruktion LPWN Rev 1 2005 Utskrift 2005 12 01 Användarmanual Vetek AB Box 79 Industriv 3 760 40 Väddö Tel 0176 208 920 Fax 0176 208 929 e mail info vetek se ww...

Page 11: ...8G tl H G W 1 14 56G tl H 7 Taiwan jin G N W _ G t l T C R L F weight Example N W 1 14 32G tl T T W 0 0 48G tl T G W 1 14 80G tl T 8 pcs T o t a l p c s C R L F pcs Example Total 0 pcs 9 C R L F Example 0 G GROSS N NET T TARE Instruktion LPWN Rev 1 2005 Utskrift 2005 12 01 Användarmanual Vetek AB Box 79 Industriv 3 760 40 Väddö Tel 0176 208 920 Fax 0176 208 929 e mail info vetek se www vetek se Si...

Page 12: ...00 kg G W 1 6025 kg N W 1 1375 kg T W 0 0180 kg G W 1 1555 kg RS 232 Connector Signal Output Instruktion LPWN Rev 1 2005 Utskrift 2005 12 01 Användarmanual Vetek AB Box 79 Industriv 3 760 40 Väddö Tel 0176 208 920 Fax 0176 208 929 e mail info vetek se www vetek se Sida 12 ...