![Vetek LPWN-1530 Manual Download Page 7](http://html.mh-extra.com/html/vetek/lpwn-1530/lpwn-1530_manual_1010068007.webp)
Instruktion-LPWN Rev 1.2005
Utskrift 2005-12-01
Användarmanual
1.
Power on while holding down MODE key. The scale will enter into function setting mode and
CAL
appears on the LCD display. Press PRINT key, the LCD display will show the scale model e.g.
3 kg
.
2.
Press MODE key once again, the LCD display will show
UON
, which means the scale enters into
weighing unit setting mode. Press PRINT key to select unit, press TARE key to enable (ON) or disable
(OFF) the weighing unit. If the display shows
ON kg
, it means the unit shown will be used;
OFF kg
means the unit shown will not be used. Press MODE key once again, the LCD display will show
Init
,which means the scale enters into initial weighing unit setting mode; press TARE key to select initial unit
setting.
3.
If the LCD display shows
Init g
, it means “g” is the initial weighing unit when power on. Available
units: kg, g, lb, lb-oz, jin, HK jin, Taiwan jin.
4.
Press MODE key once again, the LCD display will show
Fil
, which means the scale enters into
filter setting mode; press down TARE , the display will show
Fil 2
, press TARE key again to switch the
filter degree as 1, 2, 3 or 4. Level 4: Reaction is slower, effect on filtering is higher.
5.
Press MODE key once again, the LCD display will show
Auto
,which means the scale enters into
auto power off setting mode; press TARE key to select the time of auto power off. 5, 10, 30, or 60 minutes
means the lasting time value when the scale is at the non-tare zero point. In such mode, the scale will be
powered off automatically after elapse of the set time.
6.
Press MODE key once again, the LCD display will show
Ligh
, which means the scale enters into
backlight setting mode. Press TARE key to select setting, if the display shows
OFF
, it means the
backlight is off;
ON
means the backlight is on;
Auto
means auto backlight is on.
7.
Press MODE key once again, the LCD display will show
Lit
, press TARE key,
d1
will appear on the display, press TARE
key again to select zero tracking range. 0, 1, 2, 3, 4 or 5 can be
selected (the bigger the number is, the larger zero tracking range is).
8.
Press MODE key once again, the LCD display will show
BAud
, which means the scale enters into
baud rate setting mode; press TARE key to select the transmission rate of 2400, 4800, or 9600.
9.
Press MODE key once again, the LCD display will show
RS232
,which means the scale enters into
printing mode setting mode, press TARE to select
Prt Co
,
Prt St
, or
Prt Pr
(Pr means data sent
when key pressed; St means data sent automatically when stable symbol shown; Co means data sent
continuously).
10.
Press MODE key once again, the LCD display will show
, which means the scale enters into
printing mode setting mode, press TARE to select
PC
,
SH
,or
EZ
.
11.
Press MODE key once again, the LCD display will show
StAbl
, which means the scale enters into
printing stability judgement setting mode, press TARE key to select setting,
YES
means the data can be
printed out by pressing PRINT key only in the stable mode;
No
means the stable mode will not be
detected. Press MODE key to reset each function. Press ZERO key anytime when a function setting is
completed. This will make the scale return to normal weighing mode.
Vetek AB
Box 79, Industriv. 3 760 40 Väddö
Tel 0176 208 920
Fax 0176 208 929
e-mail: [email protected]
www. vetek.se
Sida
7