9.1.4 Alarming indicate
“m” is the weight
when m≥high limit, “over” lamp is lighting.
When low limit <m<high limit, “accept” lamp is lighting.
When m≤low limit, “under” lamp is lighting.
Caution:
When the display is not stabilization, no lamp lights.
9.2 Change unit
9.2.1 When it displays “units”, press [TARE], then press [SET/0~9] to choose, then press [TARE] to
confirm and exit.
Unit: H9(kg), 9(g), Lb, Lb.oz, pcs(counting). (We may not choose imperial units when calibrating).
9.2.2 counting: When it displays “pcs”, press [TARE]. Now “pcs” lamp lighting.
Longtime press [ON/zERO], it displays count and return to 0. then put the sample and press
[TARE]. Press [SET/0~9] to add 1 and press [TARE] to choose. After finishing it display pcs and pcs
lamp lighting.
Exit counting: choose a unit.
Caution: Pick quantities of Max less 30000
9.3 A-oFF(choose auto power off)
When menu displays A-oFF,press [TARE] to enter,press [SET/0~9] to choose n or y, n is not auto
power off,y is after 10 minutes in zeroing, auto-off.
9.4 Change display mode
When it displays “FILtX” (FILt1 means one step to display, FILt2 means 3-4 steps to display, FILt3
means 6-8 steps to display, FILt4 means display is fast). Press [SET/0~9] to change, press [TARE] to
confirm and exit.
9.5 Set range of auto-zero:
When it displays zero, press [TARE], now show zero x.x(x.x is the range:0.5d,1d,1.5d,2d,2.5d,3d,3.5
d,4d,4.5d,5d. d is the division value when n=3000).
Press [SET/0~9] to choose and [TARE] to confirm.
9.6 bUzzEr(set buzzle sound on or off)
When menu displays bUzzEr,press [TARE] to enter,press [SET/0~9] to choose on or off, press
[TARE] to confirm.
9.7 Change division figure
When menu displays d,press [TARE] to enter,press [SET/0~9] to choose division.
9.8 On-rA(choose range of power on)
When menu displays On-rA,press [TARE] to enter,press [SET/0~9] to choose 100, 100 is 100%
capacity. Press [TARE] to confirm.