METTLER TOLEDO 2158/2160/2162 VERTEX Floor Scale Installation and Service Manual
(2/01)
8-2
Figure 8-1: Test Weight Locations
3.
Place the test weight at the location immediately clockwise from the location at
which you got the lowest weight reading. Then adjust the trimming potentiometer for
the load cell that corresponds to the corner of the scale where the test weight is
positioned (see Figure 8-2). Make the adjustment by turning the potentiometer until
the weight reading matches the lowest reading.
Figure 8-2: Load Cell Trimming Potentiometers
4.
Proceeding clockwise, repeat the adjustment described in Step 3 for the next two
test weight locations.