background image

6

emission of light by a compound after it has absorbed a particular wavelength of
light. Under most circumstances, the emission of light will occur at a longer
wavelength than the light used to excite it. The spectrometer comes with three
LEDs (375 nm, 450 nm, and 525 nm) that serve as the excitation wavelengths.
Additional excitation LEDs can be purchased separately.

There are three general types of data collection that measure fluorescence—
fluorescence

vs.

wavelength, which produces a spectrum, fluorescence

vs.

concentration, and fluorescence

vs.

time for kinetics experiments. Once the units

have been changed to Fluorescence from the Experiment menu, follow the
instructions in the section titled Collect Data with Logger

Pro

to collect these

types of data.

These are some additional features in fluorescence mode that may improve your
data quality:

Adjusting the LED Brightness

1. Open the Spectrometer dialog box to set the LED intensity. To display this box

choose Set Up Sensors ► Spectrometer from the Experiment menu in Logger

Pro

.

2. The LED Intensity is set to 50 by default. Adjust it between 0 and 100. A

setting of 0 turns the LED off while a setting of 100 is the maximum LED
intensity.

Note:

If you adjust this value during data collection, you may want

to recalibrate or perform a manual baseline adjustment with a calculated
column.

Adjusting the Sample Time

1. Open the Spectrometer dialog box to set the Sample Time. To display this box

choose Set Up Sensors ► Spectrometer from the Experiment menu in Logger

Pro

.

2. By default, this value is set to 100 ms. The Sample Time is the amount of time

the detector is exposed to the emission light. The longer the sample time, the
greater the signal, and the longer the time it takes to collect data. 100 ms is a
good starting point for data collection. You may adjust this value while data
collection is active. If you do this, the spectrum will update in real time.

Note:

If you adjust this value during data collection, you may want to recalibrate or
perform a manual baseline adjustment with a calculated column.

Measure Emission Spectra with Logger

Pro

You may use your Spectrophotometer to measure the visible emission spectrum of
a light source such as an LED or a gas discharge tube. To do so, you will need to
purchase an optical fiber assembly (order code: VSP-FIBER).

Measure Intensity of Light Emissions

1. Insert the Spectrophotometer Optical Fiber into the Fluorescence/UV-VIS

Spectrophotometer.

2. Aim the tip of the optical fiber cable at a light source. Click

. Click

to end data collection.

Note:

The Spectrophotometer is not calibrated

for measuring intensity.

l

Option 1

The default option is to use a single 10 nm band. This measures

the average absorbance from ~5 nm on either side of the chosen
wavelength. You can change the center wavelength value by clicking on
the graph or by choosing a wavelength from the list.

l

Option 2

If you wish to use the λ max chosen by Logger

Pro

, but you

want the absorbance to be measured only at that one wavelength, change
Single 10 nm Band to Individual Wavelengths. You may then select up to
ten wavelengths to measure at the same time.

l

Option 3

If you wish to measure an average over a range of contiguous

wavelengths of your choice, change Single 10 nm Band to Individual
Wavelengths. Click

. Select boxes in the list or drag your cursor

on the graph to select up to ten contiguous wavelengths. Check Combine
Contiguous Wavelengths.

4. Click

to continue.

5. Click

. Place your first sample in the cuvette slot of the

Spectrophotometer. After the readings stabilize, click

. Enter the

concentration of the sample and click

.

6. Place your second sample in the cuvette slot. After the readings stabilize, click

. Enter the concentration of the second sample and click

.

7. Repeat Step 6 for the remaining samples. When finished, click

to end

data collection.

8. Click Linear Fit,

, to see the best fit line equation for the standard solutions.

9. If doing Beer’s law to determine the concentration of an unknown, place the

unknown sample in the cuvette holder. Choose Interpolation Calculator from
the Analyze menu. A helper box will appear, displaying the absorbance and
concentration of the unknown. Click

.

Measurement

vs.

Time (Kinetics)

1. Generate a spectrum as described above.
2. Click the Configure Spectrophotometer Data Collection button,

.

3. Select Absorbance

vs.

Time as the data-collection mode. The wavelength of

maximum absorbance will be selected. Click

to continue or select a

wavelength on the graph or in the list of wavelengths. See the previous section
for more details.

4. The default settings are 1 sample per second for 200 seconds. To change the

data-collection parameters for your experiment, choose Data Collection from
the Experiment menu and make the necessary changes. Click

.

5. Mix the reactants. Transfer ~2 mL of the reaction mixture to a cuvette and

place the cuvette in the Spectrophotometer. Click

. Click

if you

wish to end data collection early.

6. Click Curve Fit,

, to calculate a function for your data.

Measure Fluorescence with Logger

Pro

You may use your Spectrophotometer to measure the fluorescence spectrum of an
aqueous sample, such as chlorophyll, quinine, and fluorescein. Fluorescence is the

Summary of Contents for Go Direct GDX-SPEC-FUV

Page 1: ...etooth icon is solid when sensor is connected via Bluetooth wireless tech nology Go Direct Fluorescence UV VIS Spectrophotometer Order Code GDX SPEC FUV Go Direct Fluorescence UV VIS Spectrophotometer...

Page 2: ...discharge tube To do so you will need to purchase the Vernier Optical Fiber order code VSP FIBER Measure Intensity of Light Emissions Insert the Vernier Optical Fiber into the spectrometer Connecting...

Page 3: ...of your screen 1 Connect the spectrometer following the steps in the Getting Started section of this user manual 2 Open Spectral Analysis 3 Select the appropriate Emissions experiment from the listed...

Page 4: ...he Meter screen tap Mode Change the mode to Events with Entry 2 Enter the Name e g Concentration and Units e g mol L Select OK 3 A message will appear warning you to either save or discard the full sp...

Page 5: ...ments the default settings work well There are five parameters listed in the dialog box l Sample Time This is similar to the shutter speed of a camera LabQuest automatically selects the proper sample...

Page 6: ...at a light source Click Click to end data collection Note The Spectrophotometer is not calibrated for measuring intensity l Option 1 The default option is to use a single 10 nm band This measures the...

Page 7: ...e Sample Time begin with 75 ms with subsequent reductions by 20 ms to a suitable value and decrease the Samples to Average to 1 Use the Stored Emissions Files in Logger Pro Logger Pro contains a folde...

Page 8: ...this transmitter must not be co located or operating in conjunction with any other antenna or transmitter Le pr sent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio e...

Page 9: ...ensitive measurement device For best results use the cables that were provided Keep the device away from electromagnetic noise sources such as microwaves monitors electric motors and appliances Vernie...

Reviews: