background image

5

3. Follow the instructions in the dialog box to complete the calibration, and then

click

.

Collect Data with Logger

Pro

Measurement

vs.

Wavelength (Generate a Spectrum)

1. Fill a cuvette about 3/4 full of a sample of the solution to be tested. Place the

sample in the Spectrophotometer and click

. Click

to end data

collection.

2. To store the spectrum data, choose Store Latest Run from the Experiment

menu.

Measurement

vs.

Concentration (Beer’s Law Studies)

1. Generate a spectrum as described above.
2. Click the Configure Spectrophotometer Data Collection button,

.

There are three regions in this box:

l

Collection Mode

The three options for data collection are offered. If the

measurement (Absorbance in this example)

vs.

Time or

vs.

Concentration is

selected, a wavelength or wavelengths will need to be chosen.

l

Graph

The graph displays a full-spectrum analysis of the sample in the

cuvette holder. By default, the wavelength with the maximum measured value
will be selected. You may wish to choose a different wavelength by tapping on
the graph or selecting wavelength(s) from the list.

l

List of wavelength options

This column lists all the available wavelengths. It

becomes active when either the Concentration or Time mode is selected.

Configure Spectrometer data-collection dialog box

3. Select Absorbance (or %T)

vs.

Concentration as the data-collection mode. The

wavelength with the maximum value from the spectrum (λ max) will be
automatically selected. There are three options when choosing a wavelength
(or wavelengths) for subsequent measurements.

To increase the sample time, or if data collection is unusually slow, choose Set Up
Sensors ► Spectrophotometer: 1 from the Experiment menu. Set the Sample Time
(begin with 75 ms, with subsequent reductions by 20 ms) to a suitable value and
decrease the Samples to Average to 1.

Change the Settings in LabQuest

The Data Collection Screen in LabQuest lists all the settings for the device. To
display this box choose Sensors ► Data Collection from the Meter screen.

For most experiments, the default settings work well.

There are five parameters listed in the dialog box.

l

Sample Time: This is similar to the shutter speed of a camera. LabQuest
automatically selects the proper sample time during calibration.

Note:

For

emission studies, you may need to change the sample time manually.

l

Wavelength Smoothing: This is the number of adjacent readings on either side
of a given value that is used to calculate an average value.

Note:

Be careful

adjusting this parameter as it may shift your wavelength values slightly.

l

Samples to Average: This is the number of readings taken at a given
wavelength to calculate an average reading.

l

Wavelength Range: The range is determined by the type of Spectrophotometer
in use.

l

LED Intensity: This allows you to change the intensity of the excitation LED.
This setting is only available in fluorescence mode by tapping on the meter.

Using the Product with Logger

Pro

Consider using Spectral Analysis, the free app dedicated to spectroscopy. To
proceed with Logger

Pro

, connect the sensor following the steps in the Getting

Started section of this user manual.

Select the Type of Data (or Units) You Want to Measure

There are three general types of data collection that measure absorbance or
transmittance—absorbance (or %T)

vs.

wavelength, which produces a spectrum,

absorbance (or %T)

vs.

concentration for Beer’s law experiments, and absorbance

(or %T)

vs.

time for kinetics experiments.

The default data type is absorbance. If you want to measure the absorbance of a
solution, proceed directly to the Calibrate section below.

If you want to measure %T, fluorescence, or intensity, do the following:

1. Choose Change Units ► Spectrophotometer from the Experiment menu.
2. Select the unit or data type you wish to measure.

Calibrate (Optional if Measuring Fluorescence or Intensity)

1. To calibrate the Spectrophotometer, choose Calibrate ► Spectrophotometer

from the Experiment menu.

Note:

For best results, allow the

Spectrophotometer to warm up for a minimum of ten minutes.

2. Fill a cuvette about 3/4 full with distilled water (or the solvent being used in

the experiment) to serve as the blank. After the Spectrophotometer has warmed
up, place the blank cuvette in the Spectrophotometer.

Summary of Contents for Go Direct GDX-SPEC-FUV

Page 1: ...etooth icon is solid when sensor is connected via Bluetooth wireless tech nology Go Direct Fluorescence UV VIS Spectrophotometer Order Code GDX SPEC FUV Go Direct Fluorescence UV VIS Spectrophotometer...

Page 2: ...discharge tube To do so you will need to purchase the Vernier Optical Fiber order code VSP FIBER Measure Intensity of Light Emissions Insert the Vernier Optical Fiber into the spectrometer Connecting...

Page 3: ...of your screen 1 Connect the spectrometer following the steps in the Getting Started section of this user manual 2 Open Spectral Analysis 3 Select the appropriate Emissions experiment from the listed...

Page 4: ...he Meter screen tap Mode Change the mode to Events with Entry 2 Enter the Name e g Concentration and Units e g mol L Select OK 3 A message will appear warning you to either save or discard the full sp...

Page 5: ...ments the default settings work well There are five parameters listed in the dialog box l Sample Time This is similar to the shutter speed of a camera LabQuest automatically selects the proper sample...

Page 6: ...at a light source Click Click to end data collection Note The Spectrophotometer is not calibrated for measuring intensity l Option 1 The default option is to use a single 10 nm band This measures the...

Page 7: ...e Sample Time begin with 75 ms with subsequent reductions by 20 ms to a suitable value and decrease the Samples to Average to 1 Use the Stored Emissions Files in Logger Pro Logger Pro contains a folde...

Page 8: ...this transmitter must not be co located or operating in conjunction with any other antenna or transmitter Le pr sent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio e...

Page 9: ...ensitive measurement device For best results use the cables that were provided Keep the device away from electromagnetic noise sources such as microwaves monitors electric motors and appliances Vernie...

Reviews: