Inovys wheelchairs
2019-02
Adjustments
Installation manual
- 11 -
EN
NL
PL
CS
3.12
Armrest depth
1. Loosen both nuts (35) of the fixing bolts beneath
the arm cushion.
2. Remove nut and spring plunger (36).
3. Push the arm cushion into the desired position.
4. Retighten the fixing bolts (35).
5. Mount spring plunger (36) and nut (x2).
6. Verify the working of the spring plunger by moving
the arm cushion forward and backward.
7. Repeat for the second armrest. Make sure both
armrests are positioned identically.
3.13
Width between armrests
Risk of injuries
The maximum width of one armrest is limited by the position of two sockets screws. Both
need to be secured.
Adjust the width of the armrests:
1. Loosen the two socket screws (37) next to the armrest, below
the seat frame.
2. Adjust the armrest to a comfortable width (max. 50 mm
outwards).
3. Tighten the screws (37) firmly by hand.
4. Repeat for the second armrest. Make sure that the armrests
are positioned identically.
3.14
Changing battery (only for Inovys II-E)
Risk of injuries and damage
If replacement of the batteries is necessary, the complete battery pack needs to be
exchanged (IP66). Contact Vermeiren for more information.
Never try to open the battery pack.
Read and apply the instructions of the battery pack and charger to remove, install and
charge the battery. These instructions are part of the delivery.
The identification of the battery and its serial number can be found on the bottom of the
battery.
CAUTION
CAUTION
35
35
36
37