3
Warnings
* This unit must be used with the guidance of a clinician or therapist.
* Type BF equipment, Continuous Operation.
* Do not insert lead wires into a mains power supply.
* Do not immerse unit into water or any other substance.
* The unit is not protect from the ingress of water droplets from a shower of
rain if used outside the carrying case.
* Do not use this unit in the presence of a flammable anaesthetic gas mixture
and air or with Oxygen or Nitrous Oxide.
* This device is 1 x 9V PP3 Batteries operated. If using rechargeable Nickel
Metal Hydride batteries, be sure to use a CE approved battery charger.
Never connect this unit directly to a battery charger or to any other mains
powered equipment. We advise not to use Ni-Cad rechargeable batteries.
* Patient Electrodes including all skin surface electrodes, vaginal and rectal
probes are for single patient use only!
* Keep out of reach of children.
* Do not use stimulation on your facial area unless you are under strict
guidance from a qualified clinician.
* Application of electrodes near the thorax may increase the risk of cardiac
fibrillation.
* Operation in close proximity (e.g. 1m) to a shortwave or microwave
therapy equipment may produce instability in the stimulator output.
* Simultaneous connection of a patient to a high frequency surgical
equipment may result in burns at the site of the stimulator electrodes and
possible damage to the stimulator.
* No modification of this equipment is allowed!