NeuroTrac
®
MultiTENS Operation Manual
13
Programmes
MODALITIES
CON - Constant TENS
Stimulates constant frequency and pulse width on both channels.
BST - Burst TENS
Stimulates nine pulses of 200
µ
S at 150Hz repeated two times per second.
MOD - Modulated TENS
Stimulates over a six second cycle modulating frequency and / or pulse width.
P11: Frequency starts at 100Hz decreasing exponentially to 65Hz.
Pulse width starts at 200
µ
S decreasing exponentially to 100
µ
S.
P12: Frequency starts at 10Hz increasing exponentially to 90Hz.
Pulse width fixed at 200
µ
S.
HAN
3 seconds of 100Hz, followed by 3 seconds of 2 Hz.
W/R - Work / Rest Stimulation
Stimulates constant frequency and pulse width on
both channels with work / rest intervals.
Mode
Frequency
Pulse Width Work
Rest
Ramp
Up
Ramp
Down
Default
Time
TENS
P01
CON
90
200
4 Hrs
P02
CON
90
175
4 Hrs
P03
CON
80
200
4 Hrs
P04
CON
100
200
4 Hrs
P05
CON
10
200
4 Hrs
P06
CON
2
200
4 Hrs
P07
CON
30
175
4 Hrs
P08
CON
2
60
4 Hrs
P09
CON
80
60
4 Hrs
P10
BST
2 (see below)
175
4 Hrs
P11
MOD
100 / 65
200/100
4 Hrs
P12
MOD
10 / 90
200
4 Hrs
P13
HAN
100 / 2
200
4 Hrs
STIM
P14
WR
60
400
6
8
2
2
2 Hrs
P15
WR
45
200
6
6
1
1
2 Hrs
P16
WR
80
250
6
8
1
1
2 Hrs
P17
WR
60
250
10
10
1
1
2 Hrs
CUST
PC1
CON, BST, MOD,WR
PC2
CON, BST, MOD,WR
PC3
CON, BST, MOD,WR