FUNCTION SETTINGS
1. Press
[ON/OFF]
to turn on the scale, the display will show flash digits and then
--------,
press
[PRINT]
, the display will shows the
serial port
b xxx BAU
value. Press
[UNIT]
to
select among 1200, 2400, 4800 and 9600, our default setting is 9600.
Press
[ZERO]
to enter Parity mode, the display will show
P XXX PAR
, press
[UNIT]
to
select
odd, even, none
(odd stands for 7 data bits with odd parity, even stands for 7 data
bits with even parity, none stands for 8 data bit without parity)our default setting is none.
Press
[ZERO]
to enter Print mode, the display will show
n xxx Mod
, press
[UNIT]
to
select
CON,OFF,KEY,STB
(
OFF
stands for serial data output disabled,
KEY
stands for
demand mode by pressing
[PRINT],
STB stands for automatic print when scale is stable,
CON
stands for continuous print), our default setting is
STB
.
Press [
ON/OFF
] again to return to weighing mode.
2. Press
[ON/OFF]
to turn on the scale, the display will show flash digits and then
----------,
press
[UNIT]
to enter
unit selecting function
, the display will show
ON X
( X
stands for the units :g, oz, lb, dwt…tola..), press
[ZERO]
select units, press
[UNIT]
to
select
ON
(activate the selected unit) or
OFF
( inactivate the selected unit), press
[
ON/OFF
] again to return to weighing mode.
3. Switch the power to
“
ON
”
and then press
[ON/OFF]
to turn on the scale, the display
shows F-E-9 to 0 and than
“
- - - - - - -
”,
press
[PCS]
key to select
zero range
when turn
on the scale, press
[UNIT]
key to select percent (
PZR
0
,
2
,
3
,
4
,
5
,
10
,
20
,
100%FS);
Press
[ZERO]
key to set zero range, press
[UNIT]
key to select percent (
KZR
3,4,5,10,20);
Press
[ZERO]
key to set
tare range
, press
[UNIT]
key select percent (
KTR
50 %FS or
KTR
100%FS);
Press
[ZERO]
key to enter
overload protection
setting, press
[UNIT]
key to select
overload protection value (
OVR
0, 2,3,4, 5, 10,20,100%FS);
Press
[ZERO]
key to set
beep function
, press
[UNIT]
key to choose turn on or turn off
the beep,
“bZ ON”
means activate the beep,
“bZ OFF”
means inactivate the beep;
Press
[ZERO]
key to
unite the function of TARE and ZERO
, press
[UNIT]
to switch
between them.
“ZT ON”
means unite function activate,
“ZT OFF”
means unite function
inactivate.
Press
[ZERO]
key to enter
auto off function setting
,
display shows
A xxx
,
press
[UNIT]
key to select auto off time
(
3,5,10,off
)
.
Press
[ZERO]
to select
backlight,
the display will show L xxx, press
[UNIT]
to select
X
value (
X
stands for
AU,OFF,ON
),
OFF
stands for inactivate backlight function,
ON
stands
for the backlight is always make function,
AU
stands for auto backlight .
4. Press
[ZERO]
to return to weighing mode.
5
PARTS DESCRIPTION
RIGHT SIDE
AC adaptor
Socket
TOP
Display
ON/OFF
key
Turns the scale
power on/off.
ZERO
key
Sets display to zero
key
UNIT
key PCS key
Changes weighing units
AC adaptorSocket
USB
Weight
Weight cover
DISPLAY
Low battery indicator weighing unit
TARE
indicator
Negative value indicator
STABLE
indicator
when the reading is stable.
ZERO
indicator
2