SAFETY PRECAUTIONS
All safety messages are identified by the following the words "
WARNING
" and
"CAUTION"
. These words mean the following:
△
WARNING
Important information to alert you to a situation that
might cause serious injury and damage to your
property if instructions are not followed.
△
CAUTION
Important information that tells how to prevent damage
to equipment.
When using the scale/balance, the following safety precautions should always be
followed.
△
WARNING
Use only the AC adaptor with the scale/balance. Other adaptor may cause
damage. (AC adaptor is optional)
△
CAUTION
Avoid installing the scale/balance in direct sunlight, this may cause
discoloration or malfunctions.
Keep scale at room temperature, away from heat sources.
Avoid shock impact and loading over capacity. Do not keep heavy object
permanently on the platform.
If scale is not used for an extended period, keep it clean and stored in
plastic package.
Matter charged by static electricity will influence the weighing.
1
CALIBRATION
When to calibrate
1. Press [ON/OFF] to turn the power on for 15 minutes for warm up.
2.
With scale turned off, press [ON/OFF] to turn the power on.
3.
Display will show flashing digits and then “--------”,
4.
Press [ZERO], the display will show “CAL”.
5.
Press [ZERO] again , the display will show “X0000 g” where X is the
flashing digit,
6.
Press [UNIT] key to move the flashing digit to the right.
7. Press [PCS] key to increase the flashing digit. (X is the calibration
weight which can be set according to your requirement – we
recommend a minimum of 50% balance capacity).
8. Press the [ZERO] key, the display will show a series of digits known as
the A/D value, these numbers will continuously move, once the display
shows the stable indicator (please see page 16 for help locating the
stable indicator), press the [UNIT] key, the display will then show --------,
then “X00.00” or “X000.0” depending on the model and capacity.
9. Place the calibration weight on the center of the platform.
10. Press
[ZERO] key to return to weighing mode.
11. Now the Calibration is completed.
12. Press [ON/OFF] to turn off the power.
13. Press [ON/OFF] to turn on the power again.
14. Place the weight on the platform to make sure weighing is correct, if not,
repeat above steps.
6