3.0 Installation
Depending on the application of the product, it may be appropriate to position it under
an extraction hood. Ensure the extraction hood is switched on during use.
Ensure that the product is placed in such a way that it can be quickly switched off or
disconnected from the electrical supply.
Under no circumstances should any objects be placed on top of the product.
Always ensure that any vents on the top of the product are clear of any
obstruction. Always ensure all cooling vents and cooling fans (if fitted) are
clear of any obstruction.
3.3
Fitting the Chimney
l
Remove the chimney panel, (see Fig 10.)
l
Fasten the chimney to the top of the product using the screws provided, (see Fig 12.)
l
Replace the chimney panel, (see Fig 11.)