If the product is to be used in a process which could liberate hazardous gases, then it
should be installed together with a suitable fume extraction system.
Ensure that the product is placed in such a way that it can be quickly switched off or
disconnected from the electrical supply.
2.3
Setting Up
The vertical tube furnace models must have their stands assembled, the furnace
mounted on the stand and their work tubes fitted before they can be used.
Assembling the stand and mounting the furnace
Tools required: 4 mm Allen key.
The assembly of the vertical stand is as follows with reference to 2.3.1, 2.3.2 and 2.3.3.
Place the vertical stand base item (1) on the floor or bench on which it is to be
assembled.
l
Position the vertical stand upright (2) on the stand base in line with the fixing holder.
l
Secure the stand upright to the base using the 6 x M6 dome cap screws (4) and wash-
ers (3) supplied.
l
Partially screw in the top M6 screw knobs, 2 x (6) into the frame upright as shown in
2.3.3.
l
Hook the furnace onto the top screw knobs. This may take 2 people to lift the fur-
nace.
l
Fit the bottom screw knobs x 2 and tighten all four.
9
2.0 Installation