33
Operations & Maintenance Manual: Product Specifications
PRODUCT SPECIFICATIONS
COMPONENT SPECIFICATIONS
DEVICE SPECIFICATIONS
Table 8.
BladderScan Device Specifications
GENERAL SPECIFICATIONS
Bladder volume range:
0–999 mL
Accuracy:
The following accuracy specification assumes usage per instructions,
scanning a Verathon
®
tissue equivalent phantom:
Bladder Volume: ± (15% + 15 mL)
Example for a scanned volume of 160 mL:
160 mL × 15% = 24 mL
24 mL + 15 mL = 39 mL
160 mL ± 39 mL =
121–199 mL
Scan time:
Less than 5 seconds
Voice annotation interval:
10 seconds
Weight:
Less than 11 oz (309 grams)
Power:
3.7 v lithium‑ion rechargeable battery
Display:
Liquid crystal
Water resistance:
Rated at IPX1 (indicates drip‑proof, a higher than ordinary level of
protection from drips, leaks, and spills)
Expected product life
5 years
OPERATING & STORAGE SPECIFICATIONS
Operating Conditions
Temperature:
10°–40°C (50º–104ºF)
Relative humidity:
30–75% non‑condensing
Atmospheric pressure range:
70–106 kPa
Storage Conditions
Storage:
Indoor
Ambient temperature range:
‑10°–50ºC (14°–122ºF)
Atmospheric pressure range:
50–106 kPa
Relative humidity:
20–95% non‑condensing