ELGA
MEDICA 7/15 US
Operator Manual
MEDICA 7/15 US Version 2
– 10/15
Page 33
Step 1 - Start RO Sanitization Cycle
1. ENSURE that the reservoir level indication on the
graphics display is showing >40%. If display
shows >70% or 100% dispense water until
display changes to >40%.
2. PRESS the PROCESS button to stop the
process.
3. TURN OFF the electrical power supply.
4. RELIEVE residual pressure in the system by
waiting several minutes before proceeding and
then slowly opening the bleed nipple on the
capsule filter until water flow ceases. Use an
absorbent cloth to soak up the water and re-
tighten the bleed nipple.
5. DISCONNECT outlet tube from
MEDICA
at
reservoir inlet and re-direct to a suitable drain.
Step 2 - Insert a EfferSan
™ tablet into Sanitization Port
1. UNSCREW cap on sanitization port.
2. INSERT an EfferSan
™ tablet.
3. REFIT cap on sanitization port, hand tight
Step 3 - To Start the Sanitization Process
1. RESTORE the power.
2. PRESS the PROCESS button to start the
sanitization process.
3. ALLOW the sanitization cycle to continue for 20
minutes.
4. PRESS the PROCESS button to stop the
sanitization process.
5. TURN electrical supply off.
Step 4 - Return to Normal Operation
1. RECONNECT the outlet tube to the reservoir
inlet.
2. Follow steps 15 - 23 of Section 5.6 'Initial Start
up'.
8.2.2 Sanitization Procedure for the unit and docking
vessel/reservoir
The unit and docking vessel/reservoir may be sanitized to destroy
bacteria within the pipework and filters of the unit. This sanitization of
the system is recommended to be performed if high levels of bacterial
contamination are found in the product water feeding the analyser.
Should bacterial contamination of the product water be suspected,
samples should be taken by trained staff and analyzed for bacterial
counts. If the counts are abnormally high, ELGA LabWater technical
support or your local representative should be contacted for advice
and assistance.
It is recommended that only ELGA LabWater service personnel or
other fully trained staff should perform the sanitization procedure for
the unit and docking vessel/reservoir.
Sanitization Port
Insert
sanitant.