ELGA
MEDICA 7/15 US
Operator Manual
MEDICA 7/15 US Version 2
– 10/15
Page 17
Step 6 - Purity Alarm Setting
This screen is used for setting the value at which the water
purity alarm will activate.
If the
M
.cm
water purity unit setting was chosen, then the
following water purity alarm choices will be displayed.
10 M
.cm
5 M
.cm
1 M
.cm
If the
µS/cm
water purity unit setting was chosen, then the
following water purity alarm choices will be displayed.
0.1 µS/cm
0.2 µS/cm
1.0 µS/cm
Select the alarm setting required using the
button and accept
with the
button.
Note:
To reset any of the set-up parameters,
restart from the set up menu and follow
instructions from Step 1.
5.5 Setting Up Display and Replacement Timers
Turn the unit off at the power inlet module. To enter the
replacement timer set-up, press the left hand touch pad button
and at the same time turn the power on.
A graphic of the Ion-exchange cartridge, UV lamp, pre-
treatment cartridge and 0.2µm capsule filter are displayed with
hourly timer status.
CAUTION!
Before re-setting any of the cartridge
timers, ensure that the appropriate new
cartridges
have
been
installed
and
securely located correctly in the
MEDICA.
Step 1 - Ion-exchange Cartridge Replacement Timer
Setting this screen will cause the Ion-exchange cartridge timer
to reset to the preset value of 4380 hours (6 months).
Press
to jump to the next consumable or
to initiate reset.
Press
to reset timer or press to abort reset.
Press
to jump to the next consumable.
Step 2 - UV Lamp Replacement Timer
Setting this screen will cause the UV Lamp timer to reset to the
preset value of 8760 hours (1 year).
Press
to jump to the next consumable or
to initiate reset.
Press
to reset timer or press to abort reset.
Press
to jump to the next consumable.
Ion-exchange Cartridge Timer
UV Lamp Replacement Timer
Megaohm Microsiemen
Purity Alarm Setting