ELGA
MEDICA 7/15 US
Operator Manual
MEDICA 7/15 US Version 2
– 10/15
Page 11
5. INSTALLATION INSTRUCTIONS
5.1 Unpacking the
MEDICA
The following items should be supplied with your
MEDICA
:
1.
MEDICA
unit
2. Cartridge pack LC141
3. By-pass block fitted in the unit
4. Installation kit (LA637-US)
5. Operator manual
6. Mains lead
7. Black 'clip on' bend former x 4
5.2 Positioning the
MEDICA
Before commencing with installation and operation of the
MEDICA
unit, please read and observe the following points.
Environment
The unit should be installed on a flat, level surface, in a clean,
dry environment. The unit can also be wall mounted against a
vertical wall capable of supporting the weight (for this we
recommend the use of the wall mounting kit Part No LA610).
CAUTION
!
If unit is to be wall mounted, ensure it is
mounted on a substantial brick or concrete
solid wall capable of supporting the
operating
weight
of
the
system.
If
mounting the unit on the wall, use the wall
mounting kit and follow the instructions
included in the kit.
Note:
Refer to specifications for unit weights.
The unit is designed to operate safely under the following
conditions:
Indoor Use
Altitude up to 2000m
Temperature range 5 - 40°C
Maximum relative humidity 80% @ 31°C decreasing linearly
to 50% @ 40°C, non-condensating
The unit is in Installation Category II, Pollution Degree 2, as per
IEC1010-1.
Rear mounting locations points
Unit Rear Mounting Points