Cleaning and washing instructions
Cleaning of electrical and mechanical parts
The VENDLET Basic is machine washable by automatic washing
systems according to IEC 60529 (IPX6w washable).
If cleaned manually:
Clean the electrical and mechanical parts with a damp cloth
with water or a mild soap solution and then wipe the parts
with a dry cloth. You could also use a dry cloth with disinfect-
ant. All common disinfectants for normal use present in care
homes and hospitals may be used. You may also use water un-
der pressure, but the system must not be hosed down directly
with a high-pressure cleaner.
If using washing machine:
Only flat squirt nozzles are allowed. An approved setup could
be: Water Pressure 3 bar, water volume 5,61 L/min, dispersion
angle 120° flat squirt nozzle.
Mounting during washing:
Raise both side rails and turn the motor bars so that groove for
attaching the turning sheet is facing downwards.
The components should be mounted with the plug inserts
facing downwards or situated so that they are protected from
direct impact from the water nozzles. The lid covering the ca-
ble inserts on the control unit must be closed during the entire
cleaning procedure. Equally, all cables must be plugged in and
guided in such a manner, that there is no apparent strain to the
interface between the cable and its connection (for instance
between cable and control unit). It is furthermore recom-
mended that the control unit is situated as far away from the
water nozzles as possible.
Industrial high-pressure cleaning is NOT allowed.
Water:
Degree of hardness must not be more than 5° dH and no
demineralized water can be used.
Detergents:
Detergents are not allowed to contain caustic solutions. They
must not change the surface structure or adhesive properties
of the plastic and must not break down grease. pH-value must
be between 5 – 8.
Washing instructions, slide sheet
Recommended washing temperature: 60°C
Do not add fabric conditioner.
Do not tumble-dry
Do not iron
Washing instructions, turning sheet
Recommended washing temperature: 90°C
Max. 4% shrinking
Tumble-dry at normal temperatures
Iron or mangle at high temperatures
The system must not be hosed down with high
pressure cleaner
Interconnecting cables must remain plugged in
during cleaning to prevent the ingress of water
Cleaning with steam cleaner is not permitted