Press the
“Set” button for 3 seconds to access the setting mode.
Use the
“+” and “-” buttons to browse through the available menus, listed below.
Press the
“Set” button to access the selected menu. Then, use the “+” and “-” buttons to modify the value of the parameter
(when it is possible).
Press the
“Set” button to confirm.
To exit the setting mode, do not press any buttons for 5 seconds.
Text shown on
the display
Default
value
Range
Menu
’s name and description
CAL
- 0.5 °C
- 9,9 ÷ 9,9 °C
0,1 °C steps
Probe calibration
It allows to align the temperature probe reading to a
reference thermometer
ULI
50 °C
LLI
÷ 50 °C
0,1 °C steps
Setpoint upper limit
It limits the maximum value of temperature setpoint
LLI
3 °C
3 ÷
ULI
°C
0,1 °C steps
Setpoint lower limit
It limits the minimum value of temperature setpoint
CCO
---
Compressor, heating element and fan operating time
It shows compressor, heating element and fan operating
times.
Operating times are showed in hours up to 999 hours
(around 42 days). Then they are showed in days using
different combinations of the digits
’ decimal point.
CHE
CFA
UFI
NO
YES
– NO
Wi-Fi connection
It allows to enable / disable the Wi-Fi connection
OtC
---
One time code
It shows the One Time Code during the registration of
the device on the VELP Ermes cloud platform
AP
NO
YES
– NO
Access point
It allows to enable / disable the Access Point mode of
the Wi-Fi connection
rSt
NO
YES
– NO
Reset
It allows to reset all the system settings to the default
values
5. Error and warning messages
When the display shows an error message, the i
nstrument’s functions are stopped automatically.
AL1
Internal temperature too low
AL2
Internal temperature too high
AL3
Internal probe broken
AL4
Time to reach setpoint too high
Warning messages only on Ermes platform
AL5
Big internal temperature variation
To remove the error message, restart the instrument.
If the alarm persists on the display, please contact VELP Scientif
ica’s technical service department.