DVM892N
V. 03
–
19/08/2021
11
©Velleman Group nv
9.6
DIODE AND CONTINUITY
Do not conduct diode or continuity measurements on
live circuits
range
description
test condition
display reads the
approximate forward
voltage of the diode
forward DC current ± 1 mA
reversed DC voltage ± 3.0 V
built-in buzzer sounds
if resistance < 50
Ω
open-circuit voltage ± 3.0 V
Overload protection: 700 V DC or AC rms
9.7
BATTERY TEST
range
accuracy
load current
resolution
1.5 V
± (5.0 % rdg + 5 digits)
100 mA
1 mV
9 V
5 mA
10 mV
Overload protection: F500 mA/1000 V
9.8
TEMPERATURE
range
accuracy
resolution
°C
-50 to 150 °C
± (3 °C + 1 digit)
1 °C
150 to 800 °C
± (3 % + 1 digit)
NiCr-NiSi sensor
Overload protection: F500 mA/1000 V
10.
Voltage Measurement
Do not measure circuits that may contain voltages > 700 V
Use extreme caution when measuring voltages higher than 60 VDC
or 30 VAC rms.
Always place your fingers behind the protective edges of the test
probes while measuring!