Pollution
degree 2
Only nonconductive pollution occurs. Occasionally,
temporary conductivity caused by condensation is to be
expected.(home and office environments fall under this
category)
This device is only suitable for measurements in Pollution
degree class 2 environments.
operating environment:
o
temperature: 0 °C to 40 °C (32 to 104 °F)
o
humidity: < 80 % RH.
storage environment:
o
temperature: -20 °C to 60 °C (-4 °F to 140 °F)
o
humidity: < 90 % RH
9.
Specifications
overrange: yes
9.1
DC VOLTAGE
Do not measure circuits that may contain voltages
> 500 V
range
resolution
accuracy
200 mV
0.1 mV
± (1.0 % rdg + 2 digits)
2000 mV
1 mV
20 V
10 mV
200 V
100 mV
500 V
1 V
± (1.2 % rdg + 2 digits)
Overload protection: 500 V DC or AC rms
Impedance: 1M Ω
9.2
AC VOLTAGE
Do not measure circuits that may contain voltages
> 500 V
range
resolution
accuracy
200 V
100 mV
± (1.2 % rdg + 3 digits)
500 V
1 V
Average sensing, calibrated to rms of sine wave
Frequency range: 40-500 Hz
Overload protection: 500 V DC or AC rms
Impedance: 450k Ω