VELLEMAN                                                                                                   GB

3

2.1 Range Control Button

Range for AC/DC voltage, AC/DC current(µA and mA only), Resistance and
Frequency measuring can be selected manually or autorange. Push this button as
following to choose range the desired mode or range.

2.2 DATA HOLD button

When this button is pushed, the display will show the last reading and " D-H " symbol
will appear until the button is pushed again.
Data holding will be cancelled automatically when the function switch is rotated.

2.3 AC/DC Current or  /  Selecting button

Push this button to select AC or DC current measuring function when the function
switch is set at µA, mA, A positions.
Push this button to select 

 or 

 measuring when the function switch is set at 

/

position.

Power On or Function Changing

AutoRanging

Manual Ranging

Range Changed

Push button
for over
3 sec.

Push button once
"R-H"displayed

Push button for
less than 1 sec.

Summary of Contents for DVM 68

Page 1: ...DVM 68 LCD Auto Range Digital Multimeter OPERATION MANUAL GEBRUIKERSHANDLEIDING MANUEL D UTILISATEUR ...

Page 2: ...e unfamiliar with electrical circuits and testing procedures Not for commercial or industrial use 1 2 A word about safety This multimeter is designed to ensure the safest operation possible However safe operation depends on you the operator Make sure you follow these simple safety rules Never apply a voltage to the multimeter that exceeds the limits given in the specifications Never apply more tha...

Page 3: ...ies If you do not plan to use the multimeter for a month or more remove the batteries This protects the multimeter from possible leakage Disconnect the test probes before opening the multimeter Replace blown fuses only with same size and type F1 F 300mA 250V F2 F 10A 250V If any faults or abnormalities are observed do not use this device and let it check by authorised personel Never use the meter ...

Page 4: ... symbol will appear until the button is pushed again Data holding will be cancelled automatically when the function switch is rotated 2 3 AC DC Current or Selecting button Push this button to select AC or DC current measuring function when the function switch is set at µA mA A positions Push this button to select or measuring when the function switch is set at position Power On or Function Changin...

Page 5: ...nd connect test leads across the source or load under measurement 3 Read LCD display The polarity of red connection will be indicated when making a DC measurement 3 2 Measuring Current 1 Connect the black test lead to the COM jack and the red test lead to the mA Cx jack for a maximum of 300mA For a maximum of 10A move the red lead to the A jack 2 Set the function switch at µA mA or A range to be u...

Page 6: ...nge Control Button is used in this measuring function decimal points may be at incorrect positions 4 At the nF range when the capacitor to be measured is not connected to test leads the LCD may not read zero but a few counts These count have to be substracted from measuring results 3 5 Measuring Frequency 1 Connect the black test lead to the COM jack and the red test lead to the V Ω F jack 2 Set t...

Page 7: ... humidity to 80 4 1 General Max voltage between terminal and earth 1000V DC or 750V AC rms sinus Display 3 digit LCD 3260 counts max 2 3 readings sec Fuse protection µA mA range F 300mA 250V A range F 10A 250V Power Supply 9V battery Ranging method Auto Manual Polarity indication displayed Overrange indication OL displayed automatically Low Battery indication displayed Operating temperature 0 C to...

Page 8: ...load protection F 300mA fuse for µA and mA ranges F 10A fuse for A range 4 5 AC Current Range Resolution Accuracy Burden Voltage 326µA 0 1µA 0 5mV µA 3260µA 1µA 1 5 of rdg 5 digits 0 5mV µA 32 6mA 10µA 8 0mV mA 326mA 0 1mA 8 0mV mA 10A 10mA 3 0 of rdg 7 digits 0 02V A Overload protection F 300mA fuse for µA and mA ranges F 10A fuse for A range Frequency range 40Hz to 1000Hz 4 6 Resistance Range Re...

Page 9: ...al Holster 5 1 How to use the holster The holster is used to protect the meter and to make the measurement more confortable The following figures shows how to use the holster to 1 Support the meter with a standard angle fig a 2 Support the meter with a small angle using the little stand fig b 3 Hang the meter on the wall using the little stand fig c Take the little stand off from the back side of ...

Page 10: ...h a new one Fuse rarely needs replacement and blows usually as a result of the operator s error Open the case as mentioned above and take the PCB assembly out from the case Replace the blown fuse with ratings specified F1 F 300mA 250V F2 F 10A 250V Remark Before attempting to open the case be sure that test leads have been disconnected from measurement circuit to avoid electric shock ...