7 Maintenance and fault rectification
VEGASWING 53 • Transistor (PNP)
61861-EN-230208
Error
Cause
Rectification
Red signal lamp
flashes (switching
output high-im-
pedance)
Vibrating fre-
quency out of
specification
Check the vibrating element on build-
up and remove it
Buildup on the vi-
brating element
Check the vibrating element and
the sensor if there is buildup and re-
move it
Vibrating element
damaged
Check if the vibrating element is dam-
age or extremely corroded
Depending on the reason for the fault and the measures taken, the
steps described in chapter "
Setup
" must be carried out again or must
be checked for plausibility and completeness.
7.3 How to proceed if a repair is necessary
You can find an instrument return form as well as detailed information
about the procedure in the download area of our homepage. By doing
this you help us carry out the repair quickly and without having to call
back for needed information.
In case of repair, proceed as follows:
•
Print and fill out one form per instrument
•
Clean the instrument and pack it damage-proof
•
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•
Ask the agency serving you to get the address for the return ship-
ment. You can find the agency on our homepage.
Reaction after fault recti-