E
008
l
S
ensor not found
à
C
heck sensor connection
à
C
heck
HART
address of the sensor
?
E
013
l
S
ensor signals failure
,
no valid measured value
à
C
heck sensor parameter adjustment
à
R
eturn sensor for repair
?
E
016
l
E
mpty
/
full adjustment reversed
à
C
arry out a fresh adjustment
?
E
017
l
A
djustment span too small
à
C
arry out a fresh adjustment and increase the distance
between min
.
and max
.
adjustment
?
E
021
l
S
caling span too small
à
C
arry out a fresh scaling
,
increase the distance
between min
.
and max
.
scaling
.
?
E
030
l
I
nvalid measured value
à
C
heck sensor parameter adjustment
?
E
034
l
EEPROM
defective
à
C
arry out a reset
à
R
eturn instrument for repair
?
E
035
l
EEPROM CRC
-
error
à
C
arry out a reset
à
R
eturn instrument for repair
M
aintenance and fault recti
fi
cation
40
VEGASCAN
693
-
15
-
channel
HART
signal conditioning instrument
29252
-
EN
-