5 Set up with VEGADIS 82
5.1 Principle of operation and connection
The VEGADIS 82 is an external display and adjustment unit without
additional external energy.
The instrument is suitable for measured value indication and adjust-
ment of sensors with HART protocol. It can be connected at any point
to the 4 … 20 mA signal cable. A separate external energy is not
required.
4
5
3
1
2
Fig. 10: Connection of the VEGADIS 82 to the sensor, adjustment via the dis
-
play and adjustment module
1 Voltage supply/Signal output sensor
2 VEGADIS 82
3 Display and adjustment module
4 4 … 20 mA/HART signal cable
5 Sensor
5.2 Adjustment volume
Main menu:
Setup, Diagnosis, Additional adjustments, Info
Setup:
Settings, for example, for medium, application, vessel form,
adjustment, signal output
Diagnosis:
Information, for example on the instrument status, peak
value, reliability, echo curve memory as well as simulation
Additional adjustment:
False signal suppression, linearization, reset