43
13 Supplement
VEGAPULS C 11 • Two-wire 4 … 20 mA
58340-EN-200806
13 Supplement
13.1 Technical data
Note for approved instruments
The technical data in the respective safety instructions which are included in delivery are valid for
approved instruments (e.g. with Ex approval). These data can differ from the data listed herein, for
example regarding the process conditions or the voltage supply.
All approval documents can be downloaded from our homepage.
Materials and weights
Materials, wetted parts
Ʋ Antenna, process fitting
PVDF
Ʋ Counter nut
3)
PP
Ʋ Process seal
FKM
4)
Materials, non-wetted parts
Ʋ Housing
PVDF
Ʋ Cable entry seal
NBR
Ʋ Connection cable
PVC
Weight
Ʋ Instrument
0.7 kg (1.543 lbs)
Ʋ Connection cable
0.1 kg/m
Process fitting
Thread G1½, R1½, 1½ NPT
Mounting connection
Thread G1, R1, 1 NPT
Max. torque mounting boss
7 Nm (5.163 lbf ft)
Torques
Torque counter nut max.
7 Nm (5.163 lbf ft)
Input variable
Measured variable
The measured variable is the distance between the
antenna edge of the sensor and the medium surface.
The antenna edge is also the reference plane for the
measurement.