54
16 Supplement
VEGAPULS Air 41 • Autarkic device with measured value transmission via radio technology
64808-EN-210426
Packet assignment sensor status
Sensor status
Packet
2
3
4
5
6
(USA)
7
(USA)
254
Sensor function error-free
X
Sensor function error-free plus GPS infor-
mation
X
Sensor function error-free plus GPS informa-
tion (USA)
X
X
Fault
X
Error case plus GPS
X
Fault (USA)
X
X
Error case plus GPS (USA)
X
X
X
Sensor in horizontal position
X
Sensor in horizontal position plus GPS
X
Sensor in horizontal position (USA)
X
X
Sensor in horizontal position plus GPS (USA) X
X
X
Dummy required
X
Example data transmission
Packet 2, data record 02003FA31F152D2400FA09
Byte 1
Byte 2
Byte 3-6
Byte 7
Byte 8
Byte 9-10
Byte 11
0x02
0x00
0x3FA31F15
0x2D
0x24
0x00FA
0x09
Packet iden-
tifier
Namur status Measured
value
Unit
Lithium cells
Temperature
Angle of incli-
nation
2
0 = OK
1.27439
0x2D = 45 = m 36 %
25 °C
9°
Packet 5, data record 05047FFFFFFF2D24010442412A784105329B0000565409
Byte 1
Byte 2
Byte 3-6
Byte 7
Byte 8
Byte 9-
10
Byte 11-18
Byte 19-22
Byte 23
0x05
0x04
0x7FFFFFFF 0x2D
0x24
0x0104 0x42412A
784105329B
0x00005654
0x09
Packet
identi-
fier
Namur sta-
tus
Measured
value
Unit
Lithium
cells
Temper-
ature
Position
VEGA Device
status
Angle of
inclina-
tion
5
4 = fault
7FFFFFFF =
Not a Number
0x2D =
45 = m
36 %
26 °C
48.2915
8.32485
22100
9°