8. Enter new probe length and then carry out a fresh adjustment
(see "
Setup procedure, Carrying out min. adjustment - Carrying
out max. adjustment
").
1
1
1
2
2
3
A
B
C
45 mm (1.77")
100 mm (3.94")
25 mm (0.98")
50 mm (1.97")
Fig. 47: Shortening the cable probe
A Gravity weight - cable ø 4 mm
B Gravity weight - cable ø 2 mm
C Centering weight - cable ø 2 mm
1 Threaded pins
2 Thread M8 for eye-bolt
3 Fixing screw - centering weight
9.7 Software update
The following components are required to update the instrument
software:
•
Instrument
•
Voltage supply
•
Interface adapter VEGACONNECT
•
PC with PACTware
•
Current instrument software as file
You can find the current instrument software as well as detailed
information on the procedure under "www.vega.com/downloads" and
"
Software
".
Caution:
Instruments with approvals can be bound to certain software versions.
Therefore make sure that the approval is still effective after a software
update is carried out.
You can find detailed information at
www.vega.com/downloads and
"
Approvals
".
9.8 How to proceed if a repair is needed
You can find a repair form as well as detailed information on how to
proceed at www.vega.com/downloads and "
Forms and certificates
".
By doing this you help us carry out the repair quickly and without hav-
ing to call back for needed information.