4 Mounting in a standpipe
VEGAFLEX 81 • Cargo tank (crude oil)
39154-EN-130626
Mounting with gravity weight
ø 2 mm
(0.08")
ø 4 mm
(0.16")
ø 150 - 200 mm
(5.91 - 7.87")
z
y
y
z
≥ 10 mm (0.39")
3
A
B
1
2
7
4
5
6
min level
min level
Fig. 28: Mounting versions
1 Ventilation openings
2 Standpipe - vertically mounted - max. deviation 10 mm (0.4 in)
3 Gravity weight
4 Washer for centering and straining the sensor cable
5 Screw
6 Sensor cable - cannot be shortened
7 Washer for centering and straining the sensor cable
A Gravity weight with small centering disk
B Garvity weight with big centering disk - cannot be shortened
Medium
y
Water
10
mm (0.79
in)
Oil, bitumen
40
mm (3.15
in)
Solvents
40
mm (3.15
in)
Keep in mind that a min. distance must be maintained below the refer
-
ence plane - measurement is not possible here (dead band).
The value
y
states the min. lower value of the measuring range. This
value depends on the medium.
Add the height of the mounting facility
z
and the offset of the medium
y
, to calculate the total range that cannot be measured.