F
ig
.
21
:
E
xchanging the cable or rod
I
nformation
:
P
lease keep the stated torque so that the max
.
tensile strength
remains
.
T
he cable and rod of the probe can be shortened by any amount
.
1
C
able
:
loosen the three pins on the gravity weight
(
hexagon
3
)
2
C
able
:
remove the pins
3
C
able
:
P
ull the cable out of the gravity weight
4
S
horten the cable
/
rod with a cut
-
o
ff
wheel or metal saw at the
lower end
.
M
ake sure the length is correct
.
5
C
able
:
cable protrudes approx
.
40
mm
(
1
.
575
in
)
into the gravity
weight
.
6
C
able
:
F
asten the cable with three pins
,
torque
7
N
m
(
5
.
16
lbf ft
)
7
E
nter new probe length and then carry out a fresh adjustment
(
see
"
S
etup procedure
,
C
arrying out min
.
adjustment
-
C
arrying out
max
.
adjustment
"
).
S
horten cable
/
rod
40
VEGAFLEX
61
•
P
ro
fi
bus
PA
8
M
aintenance and fault recti
fi
cation
31835
-
EN
-
100426