B
asic adjustment
I
f the function
"
R
eset
"
is carried out
,
the sensor resets the values of the
following menu items to the reset values
(
see chart
):
2
)
T
he following values will be reset
:
F
unction
R
eset value
M
ax
.
adjustment
D
istance
,
upper dead zone
M
in
.
adjustment
D
istance
,
supplied sensor length
M
in
.
adjustment
-
C
able version
D
istance
,
lower dead zone
D
amping ti
0
s
L
inearisation
linear
S
ensor
-
TAG
S
ensor
D
isplay
PA
-
O
ut
1
C
hannel
(
PV
)
lin
%
A
dditional
PA
value
(
PV
)
lin
%
O
ut
-
S
cale
-
U
nit
%
PV
-
O
ut
-
S
cale
0
.
00
lin
% =
0
.
0
%
100
.
0
lin
% =
100
%
T
he values of the following menu items are
not
reset to the reset
values
(
see chart
)
with
"
R
eset
"
:
F
unction
R
eset value
S
ensor address
N
o reset
L
anguage
N
o reset
F
actory setting
L
ike basic adjustment
,
in addition
,
special parameters are reset to
default values
.
3
)
P
ointer
T
he min
.
and max
.
distance values are reset to the actual value
.
A
dditional adjustment and diagnosis options such as e
.
g
.
scaling
,
simulation or trend curve presentation are shown in the following menu
schematic
.
Y
ou will
fi
nd a detailed description of these menu items in
the operating instructions manual
"
I
ndicating and adjustment module
"
.
2
)
S
ensor
-
speci
fi
c basic adjustment
.
3
)
S
pecial parameters are parameters which are set customer
-
speci
fi
cally on
the service level with the adjustment software
PACT
ware
.
R
eset
O
ptional settings
VEGAFLEX
61
•
P
ro
fi
bus
PA
31
6
S
et up with the indicating and adjustment module
PLICSCOM
31835
-
EN
-
100426