A
linearization is necessary for all vessels in which the vessel volume
does not increase linearly with the level
-
e
.
g
.
in a horizontal cylindrical
or spherical tank
-
and the indication or output of the volume is
required
.
C
orresponding linearization curves are preprogrammed for
these vessels
.
T
hey represent the correlation between the level
percentage and vessel volume
.
B
y activating the appropriate curve
,
the volume percentage of the vessel is displayed correctly
.
I
f the
volume should not be displayed in percent but e
.
g
.
in l or kg
,
a scaling
can be also set in the menu item
"
D
isplay
"
.
L
inearisation curve
linear
E
nter the requested parameter via the appropriate keys
,
save your
settings and jump to the next menu item with the
[-
>
]
key
.
C
aution
:
N
ote the following
,
if
VEGACAL
62
is used as part of an over
fi
ll
protection system according to
WHG
:
I
f a linearisation curve is selected
,
the measuring signal is no longer
compulsorily linear proportional to the level
.
T
his must be taken into
consideration by the user
,
particularly when adjusting the switching
point on the level switch
.
I
n this menu item you can enter an unambiguous designation for the
sensor
,
e
.
g
.
the measurement loop name or the tank or product
designation
.
I
n digital systems and in the documentation of larger
plants
,
a singular designation should be entered for exact identi
fi
cation
of individual measuring sites
.
S
ensor
-
TAG
S
ensor
W
ith this menu item
,
the
B
asic adjustment is
fi
nished and you can now
jump to the main menu with the
[
ESC
]
key
.
A
dditional adjustment and diagnosis options such as e
.
g
.
scaling
,
simulation or trend curve presentation are shown in the following menu
schematic
.
Y
ou will
fi
nd a detailed description of these menu items in
the operating instructions manual
"
I
ndicating and adjustment module
"
.
B
asic adjustment
I
f the function
"
R
eset
"
is carried out
,
the sensor resets all settings to
default
.
T
he following values will be reset
:
L
inearisation curve
S
ensor
-
TAG
O
ptional settings
R
eset
30
VEGACAL
62
•
P
ro
fi
bus
PA
6
S
et up with the indicating and adjustment module
PLICSCOM
30025
-
EN
-
090603
Summary of Contents for VEGACAL 62
Page 1: ...Operating Instructions VEGACAL 62 Pro bus PA Document ID 30025 Capacitive...
Page 14: ...Fig 5 In owing medium 14 VEGACAL 62 Pro bus PA 4 Mounting 30025 EN 090603...
Page 57: ...VEGACAL 62 Pro bus PA 57 10 Supplement 30025 EN 090603...
Page 58: ...58 VEGACAL 62 Pro bus PA 10 Supplement 30025 EN 090603...
Page 59: ...VEGACAL 62 Pro bus PA 59 10 Supplement 30025 EN 090603...