T
hese values can be helpful if the electronics has to be exchanged
.
%
V
alue
E
mpty adjustment
F
ull adjustment
T
ab
.
1
:
A
djustment protocol
T
ip
:
F
or min
.
adjustment the vessel should be as empty as possible
,
and
for max
.
adjustment
,
as full as possible
.
I
f the vessel is already full
,
start with max
.
adjustment
.
N
ote
:
I
f possible
,
the vessel should be as empty as possible for min
.
adjustment
.
P
roceed as follows
:
1
M
ove from the measured value display to the main menu by
pushing
[
OK
]
.
▶
B
asic adjustment
D
isplay
D
iagnostics
S
ervice
I
nfo
2
S
elect the menu item
"
B
asic adjustment
"
with
[-
>
]
and con
fi
rm with
[
OK
]
.
N
ow the menu item
"
M
in
.
adjustment
"
is displayed
.
M
in
.
adjustment
0
.
00
%
=
0
.
0
p
F
54
.
5
p
F
3
P
repare the adjustment value for editing with
[
OK
]
.
M
ove to the
selection window with
[
OK
]
.
M
in
.
adjustment
A
ccept current meas
-
ured value
?
A
ccept
?
E
dit
?
4
A
ccept the current measured value or move to the editing window
with
"
E
dit
"
.
T
o edit
,
set the cursor to the requested position with
[-
>
]
.
S
et the requested
%
value with
[
+
]
and save with
[
OK
]
.
T
he
cursor jumps to the capacitance value
.
5
E
nter the current capacitance value in p
F
(
displayed below
)
for the
empty vessel corresponding to the percentage value
.
28
VEGACAL
62
•
P
ro
fi
bus
PA
6
S
et up with the indicating and adjustment module
PLICSCOM
30025
-
EN
-
090603
Summary of Contents for VEGACAL 62
Page 1: ...Operating Instructions VEGACAL 62 Pro bus PA Document ID 30025 Capacitive...
Page 14: ...Fig 5 In owing medium 14 VEGACAL 62 Pro bus PA 4 Mounting 30025 EN 090603...
Page 57: ...VEGACAL 62 Pro bus PA 57 10 Supplement 30025 EN 090603...
Page 58: ...58 VEGACAL 62 Pro bus PA 10 Supplement 30025 EN 090603...
Page 59: ...VEGACAL 62 Pro bus PA 59 10 Supplement 30025 EN 090603...