15
5 Set up with the display and adjustment module
VEGABAR 86 • Foundation Fieldbus
46323-EN-210430
Diagnosis:
Information, for example, of device status, peak value,
simulation
Additional adjustments:
date/time, reset, copy function
Info:
Instrument name, hardware and software version, factory cali-
bration date, device ID, sensor features
Note:
For optimum setting of the measuring point, the individual submenu
items in the main menu item "
Setup
" should be selected one after
the other and provided with the correct parameters. If possible, go
through the items in the given sequence.
The procedure is described below.
The following submenu points are available:
The submenu points are described below.
5.4 Menu overview
The following tables show the adjustment menu of the instrument.
Depending on the instrument version or application, all menu items
may not be available or some may be differently assigned.
Setup
Menu item
Parameter
Default value
Application
Application
Level
Secondary sensor for electronic differen-
tial pressure
Deactivated
Units
Adjustment unit (m, bar, Pa, psi … user-
defined)
mbar (with nominal measuring range
≤ 400 mbar)
bar (with nominal measuring ranges
≥ 1 bar)
Temperature unit (°C, °F)
°C
Position correction
0.00 bar
Adjustment
Zero/Min. adjustment
0.00 bar
0.00 %
Span/Max. adjustment
Nominal measuring range in bar
100.00 %
Damping
Integration time
1 s
Linearisation
Linear, cylindrical tank, … user-defined
Linear
Lock adjustment
Blocked, released
Released